FreeStyle Flash Blood Glucose Monitors System. Meter made in China; Distributed by TheraSense, Alameda, CA The FreeStyle Flash meter should only be used with FreeStyle Test Strips and FreeStyle Control Solution. Using other brands of test strips and control solutions with the FreeStyle Meter can product inaccurate results. Intended to monitor blood glucose from samples taken from the body. IVD use only,

Publicado el 26 de marzo de 2014 · Inicio del retiro: 19 de febrero de 2014

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Abbott Diabetes Care has identified through internal testing and investigation that all non-applied voltage legacy meters have the potential to produce out of range control solution results and/or erroneously low blood glucose results when used in conjunction with Free Style test strips lots within expiry.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Abbott Diabetes Care, Inc.
Distribuidor
Worldwide Distribution: US (nationwide) and countries of: Austria, Canada, Germany, Israel, Netherlands, Norway, Sweden, Switzerland, United Kingdom.
Lote
Catalog Number 17002 NDC 99073-0170-02. All serial numbers
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States