FDAFDA · Alimentos y medicamentos·N.º oficial Z-1211-2014

FreeStyle Blood Glucose Monitors System. For in vitro diagnostic use only. Store at room temperature below 86 degrees F (30 degrees C). Use test strips only within the system operating temperature range as outlined in your Owner's Booklet. Meter made in China; Distributed by Abbott Diabetes Care, Alameda, CA The FreeStyle mete should only be used with FreeStyle Test Strips and FreeStyle Control Solution. Using other brands of test strips and control solutions with the FreeStyle Mete can product inaccurate results. The FreeStyle Glucose meter is incorporated into the OmniPod insulin Management system (manufactured by Insulet Corporation, Bedford, MA and is intended for subcutaneous delivery of insulin. Intended to monitor blood glucose from samples taken from the body. IVD use only,

Publicado el 26 de marzo de 2014 · Inicio del retiro: 19 de febrero de 2014

Resumen

FDA registra el retiro Z-1211-2014 de dispositivos médicos de Clase I, publicado el 26 de marzo de 2014 e iniciado el 19 de febrero de 2014. Empresa responsable: Abbott Diabetes Care, Inc.; Estado: Finalizado; Cantidad reportada: not provided. Device has not been shipped in the last 5 years.; Lote: Part number CAT11001-0; all serial numbers. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Abbott Diabetes Care has identified through internal testing and investigation that all non-applied voltage legacy meters have the potential to produce out of range control solution results and/or erroneously low blood glucose results when used in conjunction with Free Style test strips lots within expiry.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Abbott Diabetes Care, Inc.
Distribuidor
Worldwide Distribution: US (nationwide) and countries of: Austria, Canada, Germany, Israel, Netherlands, Norway, Sweden, Switzerland, United Kingdom.
Lote
Part number CAT11001-0; all serial numbers
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿FreeStyle Blood Glucose Monitors System. For in vitro diagnostic use only. Store at room temperature below 86 degrees F (30 degrees C). Use test strips only within the system operating temperature range as outlined in your Owner's Booklet. Meter made in China; Distributed by Abbott Diabetes Care, Alameda, CA The FreeStyle mete should only be used with FreeStyle Test Strips and FreeStyle Control Solution. Using other brands of test strips and control solutions with the FreeStyle Mete can product inaccurate results. The FreeStyle Glucose meter is incorporated into the OmniPod insulin Management system (manufactured by Insulet Corporation, Bedford, MA and is intended for subcutaneous delivery of insulin. Intended to monitor blood glucose from samples taken from the body. IVD use only, fue retirado del mercado?

Sí. FDA publicó el retiro Z-1211-2014 el 26 de marzo de 2014. Peligro reportado: Abbott Diabetes Care has identified through internal testing and investigation that all non-applied voltage legacy meters have the potential to produce out of range control solution results and/or erroneously low blood glucose results when used in conjunction with Free Style test strips lots within expiry.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter