Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1252-2014

Natural-Knee¿ II Unicompartmental Knee System: UNI ALL-POLY TIB SZ 1-7mm UNI ALL-POLY TIB SZ 1-9mm

Publicado el 26 de marzo de 2014 · Inicio del retiro: 25 de febrero de 2014

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.

Qué hacer

Voluntary: Firm initiated

Detalles del retiro

Fabricante
Zimmer, Inc.
Distribuidor
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
Lote
Item Number 613001007 613001009
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States