Obsidio Conformable Embolic, Material Numbers (UPN): a) M0013972001010, b) M0013972101010
Publicado el 3 de abril de 2024 · Inicio del retiro: 20 de febrero de 2024
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
An investigation determined that delivery of the Obsidio embolic using the aliquot technique for lower gastrointestinal bleeding embolization poses a high risk of bowel ischemia. The most serious and the most common adverse health consequence, reasonably foreseeable to occur, is the need to perform major surgery such as bowel resection and/or diverting colostomy. Therefore, Boston Scientific does not recommend that the aliquot technique be used to deliver the Obsidio device for lower GI bleed embolization procedures.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Boston Scientific Corporation
- Distribuidor
- US
- Lote
- a) M0013972001010, UDI/DI 00191506039332, ALL LOT CODES b) M0013972101010, UDI/DI 00191506043124, ALL LOT CODES
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?
Recalert vigila todo tu catálogo contra retiros como este, automáticamente.