Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1491-2014

The CEP 8 SpectrumGreen (SG) ASR Probe Kit, 20ul (list 06J37-018) is an Analyte Specific Reagent (ASR) containing a single vial of CEP 8 spectrum green probe and a Certificate of Analysis (COA). Chromosome enumeration probes (CEP) are chromosome-specific FISH probes that hybridize to highly repetitive human satellite DNA sequences, usually located near centromeres. CEP signals enable the identification and enumeration of human chromosomes in interphase and metaphase cells from fresh and archived samples. The label for CEP 8 SG ASR Probe Kit indicates it is an Analyte Specific Reagent, and analytical and performance characteristics are not established.

Publicado el 30 de abril de 2014 · Inicio del retiro: 28 de marzo de 2013

Resumen

FDA registra el retiro Z-1491-2014 de dispositivos médicos de Clase III, publicado el 30 de abril de 2014 e iniciado el 28 de marzo de 2013. Empresa responsable: Abbott Molecular; Estado: Finalizado; Cantidad reportada: 20 kits; Lote: List Number: 06J37-018; Lot Number: 440023. La Clase III indica que no es probable que el producto cause consecuencias adversas para la salud.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

20 kits of CEP 8 SG ASR Probe (list 06J37-018, lot 440023), contained Vysis CEP 4 (Alpha Satellite) SG Probe, 20u, (part 32-112004 lot 438814), instead of the expected CEP 8 probe.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Abbott Molecular
Distribuidor
Nationwide (CA, CO, CT, FL, HI, IL, IN, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, TN, TX, VA, WA, WI); Canada
Lote
List Number: 06J37-018; Lot Number: 440023
Clasificación de riesgo
Class III
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿The CEP 8 SpectrumGreen (SG) ASR Probe Kit, 20ul (list 06J37-018) is an Analyte Specific Reagent (ASR) containing a single vial of CEP 8 spectrum green probe and a Certificate of Analysis (COA). Chromosome enumeration probes (CEP) are chromosome-specific FISH probes that hybridize to highly repetitive human satellite DNA sequences, usually located near centromeres. CEP signals enable the identification and enumeration of human chromosomes in interphase and metaphase cells from fresh and archived samples. The label for CEP 8 SG ASR Probe Kit indicates it is an Analyte Specific Reagent, and analytical and performance characteristics are not established. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1491-2014 el 30 de abril de 2014. Peligro reportado: 20 kits of CEP 8 SG ASR Probe (list 06J37-018, lot 440023), contained Vysis CEP 4 (Alpha Satellite) SG Probe, 20u, (part 32-112004 lot 438814), instead of the expected CEP 8 probe.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter