Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1514-2020

IMMULITE 2000 Systems E2 Estradiol Catalog #: US: L2KE22(D) (200 tests) and L2KE26(D) (600 tests) and OUS: L2KE22 (200 tests), L2KE26 (600 tests) - Product Usage: For in vitro diagnostic use with the IMMULITE systems Analyzers for the quantitative measurement of estradiol in serum, as an aid in the diagnosis and treatment of various sexual disorders.

Publicado el 25 de marzo de 2020 · Inicio del retiro: 5 de febrero de 2020

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Due to high discordant Estradiol results which.potentially affects all patient populations, A falsely elevated estradiol level could lead a clinician to misinterpret a patient as pre-menopausal when truly post- menopausal. This may lead to delayed initiation of a potentially beneficial drug and/or administration of an unnecessary drug in the treatment for hormone receptor positive advanced or metastatic breast cancer.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Siemens Healthcare Diagnostics, Inc.
Distribuidor
Worldwide distribution - US Nationwide in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, IN, KY, LA, MA, MD, MI, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, RI. SC, TN, TX, UT, VA, WA, WI, WY, and countries of AE, AF, AL, AO, AR, AT, AU, AZ, BD, BE, BO, BG, BH, BR, CA, CH, CO, CL, CN, CO, CR, CY, DE, DO, DZ, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, IE, IN, IR, IS, IT, IQ, JO, KE, KR, LB, LT, LV, LY, ME, MK, MX, NL, NO, PA, PE, PH, PK, PL, PS, PT, PY, QA, RO, RU, SA, SE, SK, SL, SV, SY, TH, TJ, TM, TN, TR, TW, UA, UY, UZ, VE, VN, XK, XS, ZA
Lote
Manufacturing Lot #: All kit lots beginning 501 and above. SMN & (UDI) for L2KE22(D) US: 10702833 (00630414947907), L2KE26(D) 10702834 (00630414947914). SMN & (UDI) for L2KE22 OUS: 10381178 (00630414966106), L2KE26: 10381177 (00630414966007).
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta