Owner's Booklets and Instructions for Use that are used with the following blood glucose measurement devices. Description/Item: TRUE METRIX PRO Meter Only/RE4H01P-40, TRUE METRIX PRO Starter Kit/RE4H01P-43, PROCURE TRUE METRIX PRO Meter Only/RE4212P-00, Henry Schein TRUE METRIX PRO Meter Only/RE4099P-40, Henry Schein TRUE METRIX PRO Kit/RE4099P-43, McKESSON TRUE METRIX PRO Meter Only/RE4051P-00, Moore Medical TRUE METRIX PRO Meter Only/RE4068P-01.
Publicado el 18 de marzo de 2026 · Inicio del retiro: 6 de febrero de 2026
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Trividia Health, Inc.
- Distribuidor
- Worldwide distribution - US Nationwide including Puerto Rico and the countries of Mexico, United Kingdom, Jamaica, Trinidad & Tobago, and Australia.
- Lote
- Device Item/UDI-DI: RE4H01P-40/21292006099, RE4H01P-43/21292006105, RE4212P-00/10840330705681, RE4099P-40/21292014919, RE4099P-43/21292008673, RE4051P-00/20612479197217, RE4068P-01/607415261097. Owner's Booklet or IFU Description/Part and Revision: TRUE METRIX PRO Owners Booklet/RE4TVHP03 Rev 56 and prior, McKesson Med Surg TRUE METRIX PRO Owners Booklet/ RE4SUNP03 Rev 56 and prior, Device serial numbers: Prefix "TP" followed by a seven-digit numeric sequence.
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States