Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1564-2016

Magellan Robotic Catheter Accessory Kits, Model number MACC 50 and MACC80. Cardiology: The Hansen Medical Magellan System is a robotic control system that enables positioning and visualization of a Magellan Catheter and an off-the-shelf guide wire while allowing a physician to remain seated and away from the x-ray radiation source. The Accessory Kits contain disposable components that keep the Magellan Catheters and 3rd-party guide wires performing smoothly during system operation.

Publicado el 4 de mayo de 2016 · Inicio del retiro: 25 de marzo de 2016

Resumen

FDA registra el retiro Z-1564-2016 de dispositivos médicos de Clase II, publicado el 4 de mayo de 2016 e iniciado el 25 de marzo de 2016. Empresa responsable: Hansen Medical Inc; Estado: Finalizado; Cantidad reportada: 8 kits; Lote: Lot number 20151117. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Firm's quarterly dose audit testing showed that one batch was radiated below the augmented dose and therefore did not meet the augmented dose of 28.4 kGy.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Hansen Medical Inc
Distribuidor
Nationwide Distribution to OH, TX. FL, CA and GA.
Lote
Lot number 20151117
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Magellan Robotic Catheter Accessory Kits, Model number MACC 50 and MACC80. Cardiology: The Hansen Medical Magellan System is a robotic control system that enables positioning and visualization of a Magellan Catheter and an off-the-shelf guide wire while allowing a physician to remain seated and away from the x-ray radiation source. The Accessory Kits contain disposable components that keep the Magellan Catheters and 3rd-party guide wires performing smoothly during system operation. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1564-2016 el 4 de mayo de 2016. Peligro reportado: Firm's quarterly dose audit testing showed that one batch was radiated below the augmented dose and therefore did not meet the augmented dose of 28.4 kGy.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter