Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1621-2013

Bigliani/Flatow¿ The Complete Solution Fukuda Retractors Small Nonsterile 00-4305-019-30. Used in retracting the humeral shaft posteriorly and exposing the entire glenoid surface.

Publicado el 10 de julio de 2013 · Inicio del retiro: 23 de mayo de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

This is a recall expansion from the 2010 recall which now includes seven additional lots that were not previously included. Zimmer, Inc. is initiating a recall of small and large Fukuda-type retractors due to fractures occurring at the grooves around the perimeter of the blades while being used during surgery.

Qué hacer

Voluntary: Firm initiated · Telephone

Detalles del retiro

Fabricante
Zimmer, Inc.
Distribuidor
Worldwide Distribution-USA (nationwide) including the states of NY, MI, MN, OH, IN, IL, VA, MD, PA, WA, KY, CT, RI, NH, TX, MT, and CA and the countries of Chile, China, South Korea, and Australia.
Lote
Item Number 00-4305-019-30 with Lot Numbers 61557761, 61591732, 61612275, 61614764, and 77000186. 2010 Market Withdrawal part numbers: Item #:430501930, Lot#s: 60795817, 60834923, 60916795, 60997103, 61017041, 61038697, 61075510, 61128521, 61186400, 61199809, 61253702, 61392455, 61430126
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta