MAQUET Getinge Group Mobile Operating Table YUNO OTN GA 1433.01 EN 05. The YUNO OTN is used to support and position patients for surgical procedures immediately before, during and after the surgical phase. The table top is radio translucent and enables intraoperative use of X-ray equipment.
Publicado el 10 de julio de 2013 · Inicio del retiro: 26 de abril de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
MAQUET has identified a potential issue with the extension device handle (part number 1433.62A1) and will make enhancements to the YUNO table (part number 1433.01F0). Identification of the Extension device issue: By fully actuating the lever of the extension device handle, the operator's skin can enter a gap between the handle lever and a recess in the handle bar. When the lever is released,
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Maquet Cardiovascular Us Sales, Llc
- Distribuidor
- Worldwide Distribution - USA including Arizona, Georgia, North Dakota, South Carolina, Texas and Tennessee, Internationally to United Arab Emirates, Austria, Brazil, Canada, Switzerland, China, Germany, Spain, France, United Kingdom, Italy, Japan, Singapore, Thailand, Turkey, Serbia, and South Africa.
- Lote
- YUNO OTN Mobile Operating Table: 1433.01F0 Extension Device: 1433.62A1
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States