FDAFDA · Alimentos y medicamentos·N.º oficial Z-1716-2024
ZPLP(Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 6 Holes, 106 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 47-2357-018-06
Publicado el 8 de mayo de 2024 · Inicio del retiro: 20 de marzo de 2024
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
Qué hacer
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Detalles del retiro
- Fabricante
- Zimmer, Inc.
- Distribuidor
- Worldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.
- Lote
- UDI: (01)00889024088573(17)290510(10)64370380 (01)00889024088573(17)290517(10)64370382 (01)00889024088573(17)290607(10)64423329 (01)00889024088573(17)290811(10)64475643 (01)00889024088573(17)300225(10)64661551 (01)00889024088573(17)300307(10)64661552 (01)00889024088573(17)300511(10)64745739 (01)00889024088573(17)300506(10)64745742 (01)00889024088573(17)300506(10)64745744 (01)00889024088573(17)300506(10)64745745 (01)00889024088573(17)321209(10)65635924 (01)00889024088573(17)321207(10)65645147 (01)00889024088573(17)330215(10)65755766 (01)00889024088573(17)330430(10)65997131 (01)00889024088573(17)330428(10)65997134 (01)00889024088573(17)330430(10)66023658 (01)00889024088573(17)330730(10)66166098 (01)00889024088573(17)330811(10)66201200 (01)00889024088573(17)330822(10)66233083 (01)00889024088573(17)330822(10)66233084 Lot Numbers: 64370380 64370382 64423329 64475643 64661551 64661552 64745739 64745742 64745744 64745745 65635924 65645147 65755766 65997131 65997134 66023658 66166098 66201200 66233083 66233084
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States