CDI Blood Parameter Monitoring System 500 with Arterial and Venous blood parameter modules. Intended for use during cardiopulmonary bypass procedures when continuous monitoring of arterial and/or venous pH, PCO2, PO2, K+, oxygen saturation, hematocrit, hemoglobin, and temperature are desired.
Publicado el 24 de julio de 2013 · Inicio del retiro: 27 de febrero de 2013
Resumen
FDA registra el retiro Z-1732-2013 de dispositivos médicos de Clase II, publicado el 24 de julio de 2013 e iniciado el 27 de febrero de 2013. Empresa responsable: Terumo Cardiovascular Systems Corporation; Estado: Finalizado; Cantidad reportada: 3972 since 4/26/12; Lote: Catalog number: 500AV* and serial numbers: 0024, 0033, 0040, 0051, 1192, 1193, 1355,1356, 1420, 1423, 1464, 1465, 1467, 1468,1483,1504,1521-1524,1526,1527,1529,1530, 1602-1604, 1606, 1662, 2023, 2035, 2102-2107, 2800, 2901,2902, 3131-3135, 3261, 3262, 4132, 5330-5333, 5338-5340, 5578, and 5579. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for Hematocrit/Saturation calculations.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Terumo Cardiovascular Systems Corporation
- Distribuidor
- Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV and the countries of Jordan, CANADA, AUSTRALIA, Japan ,Hong Kong ,UNITED ARAB EMIRATES (UAE), BELGIUM, South Korea, Malaysia, Mexico, Philippines, Singapore, Taiwan, and Thailand.
- Lote
- Catalog number: 500AV* and serial numbers: 0024, 0033, 0040, 0051, 1192, 1193, 1355,1356, 1420, 1423, 1464, 1465, 1467, 1468,1483,1504,1521-1524,1526,1527,1529,1530, 1602-1604, 1606, 1662, 2023, 2035, 2102-2107, 2800, 2901,2902, 3131-3135, 3261, 3262, 4132, 5330-5333, 5338-5340, 5578, and 5579.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDACDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM2 de julio de 2025
- FDACardiovascular Procedure Kits that include High Crack Check Valve PN 205307A Part Number: 6375 16010 6297410 de abril de 2024
- FDA(1) Capiox FX25 (West) Advance Oxygenator- with left port, 4-liter Reservoir, XC modified to Advance, Sterile23 de diciembre de 2020
- FDATerumo Capiox NX19 Oxygenator With Reservoir (East Orientation) P/N: 3CX*NX19RE7 de agosto de 2019
- FDACDI Blood Parameter Monitoring System 500 with Arterial blood parameter modules and Hematocrit/Saturation24 de julio de 2013
- FDACDI Blood Parameter Monitoring System 500 with Arterial blood parameter module. Intended for use during24 de julio de 2013
Preguntas sobre este retiro
¿CDI Blood Parameter Monitoring System 500 with Arterial and Venous blood parameter modules. Intended for use during cardiopulmonary bypass procedures when continuous monitoring of arterial and/or venous pH, PCO2, PO2, K+, oxygen saturation, hematocrit, hemoglobin, and temperature are desired. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1732-2013 el 24 de julio de 2013. Peligro reportado: Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for Hematocrit/Saturation calculations.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter