Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1737-2013

KRD1 PedFuse Rods, Model: IM80100-13 Rev B Product Usage: Pedicle rods are cylindrical rods which act as a linking mechanism between pedicle screws. The rod sits with the pedicle screw tulip and they are locked into place with a set screw or locking cap. The rod may be placed along several pedicle screws. Component rods are used during implantation of pedicle screw system devices intended to provide immobilization and stabilization of spinal segments of skeletally mature patients.

Publicado el 24 de julio de 2013 · Inicio del retiro: 30 de diciembre de 2011

Resumen

FDA registra el retiro Z-1737-2013 de dispositivos médicos de Clase II, publicado el 24 de julio de 2013 e iniciado el 30 de diciembre de 2011. Empresa responsable: SpineFrontier, Inc.; Estado: Finalizado; Cantidad reportada: 2 rods; Lote: Lot Number: 002876-007R. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

One of the rods in each of two test constructs did not meet fatigue requirement as outlined in ASTM F1717-04 dynamic compression testing 5 million cycles.

Qué hacer

Voluntary: Firm initiated · Telephone

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
SpineFrontier, Inc.
Distribuidor
USA Nationwide Distribution including the state of TX
Lote
Lot Number: 002876-007R
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿KRD1 PedFuse Rods, Model: IM80100-13 Rev B Product Usage: Pedicle rods are cylindrical rods which act as a linking mechanism between pedicle screws. The rod sits with the pedicle screw tulip and they are locked into place with a set screw or locking cap. The rod may be placed along several pedicle screws. Component rods are used during implantation of pedicle screw system devices intended to provide immobilization and stabilization of spinal segments of skeletally mature patients. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1737-2013 el 24 de julio de 2013. Peligro reportado: One of the rods in each of two test constructs did not meet fatigue requirement as outlined in ASTM F1717-04 dynamic compression testing 5 million cycles.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Telephone