RayStation software version 4.0.3.4 RayStation is a software system designed for treatment planning and analysis of radiation therapy.

Publicado el 25 de junio de 2014 · Inicio del retiro: 25 de abril de 2014

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

This notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of an ROI can be wrong. This affects all dose-volume properties for such ROIs, including DVH, dose statistics, clinical goals and constraints or objective functions. Furthermore, if material override ROIs are used, or if the external is not represented by contours, the bug can trigger an error in the 3D dose distribution.

Qué hacer

Voluntary: Firm initiated · E-Mail

Detalles del retiro

Fabricante
RAYSEARCH LABORATORIES AB
Distribuidor
Worldwide Distribution-USA (nationwide) including the states of AZ, CA, FL, IL, MA, MD, MO, MS, NJ, NV, NY, OH, PA, TN, VA, and WA, and the countries of Australia, Belgium, Germany, France, Italy, New Zealand, Republic of Korea, Switzerland, Spain, and United Kingdom.
Lote
Software version 4.0.3.4
Clasificación de riesgo
Class II
Área (FDA)
device
País
Sweden