Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1819-2016

Various trauma and sports medicine instruments and implants. Rod, fixation, intramedullary and accessories.

Publicado el 1 de junio de 2016 · Inicio del retiro: 1 de abril de 2016

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an infection.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Biomet, Inc.
Distribuidor
Worldwide Distribution - US (Nationwide) and globally to Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecaudor, The Netherlands, Japan, Korea, Malaysia, Mexico, Peru.
Lote
Part #'s 28234 28338 28811 28813 28821 29260 29822 805065080 14-400428 14-400436 14-405032 14-405040 14-440118 14-440121 14-441280 14-440066S Lot #'s 590970 591080 591090 665210 692040 692250 692280 692290 695900 610470 665230 665240 692440 692450 692480 692620 692660 692670 692690 721680 721720 721730 729080 722730 621840 621860 678300 664740 664820 664870 664880 503760 503770 693840 476230
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta