216 Wilson Frame Pads, Model # 5319-37; Used with Radiolucent Wilson Frames, model numbers 5319G and 5321G. General and Plastic Surgery: These pads are used as a component of the Wilson Frames, which support the torso of a prone patient during spine surgery. The pads are supplied with the frames and as replacements.

Publicado el 26 de abril de 2017 · Inicio del retiro: 28 de marzo de 2017

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Mizuho has discovered that the 5319-37 Wilson Frame Pads have a design deficiency in the construction of their mounting sleeves that may allow the patient to contact the underlying carbon fiber structure of the Wilson frame and could lead to a pressure injury.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Mizuho OSI
Distribuidor
United States, Canada, Japan, Australia, New Zealand, Great Britain, Sweden, Spain, Germany, Switzerland, Kenya, Saudi Arabia, Hong Kong, South Korea.
Lote
All Lots: Manufacturing dates: December 2016 through March 2017; GTIN:00842430100192
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States