syngo Lab Data Manager, SMN 10800057, 10803189, and 10804573.

Publicado el 9 de julio de 2014 · Inicio del retiro: 20 de mayo de 2014

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Certain configuration parameters may cause the reference ranges to not hold results as expected and release them to the Laboratory Information System (LIS) that should have been held for manual review.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Siemens Healthcare Diagnostics, Inc.
Distribuidor
Worldwide Distribution. US Nationwide, Canada, Australia, Belgium, Czech Republic, Germany, Denmark, Finland, France, United Kingdom, Italy, Netherlands, Portugal, Slovakia, Singapore, Japan, and Spain.
Lote
syngo Lab Data Manager Application Software versions VA11B, VA12A, and VA12B
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States