Infinite F500, in vitro diagnostic. Product Usage: The Tecan Infinite F500 is a multifunctional microplate reader with injector option. The Infinite F500 provides high performance for the vast majority of todays microplate applications and research and is robotic compatible. The Infinite F500 is intended as a general purpose laboratory instrument for professional use, supporting common microplate conforming to the ANS/SBS standards.
Publicado el 8 de julio de 2015 · Inicio del retiro: 29 de mayo de 2015
Resumen
FDA registra el retiro Z-1967-2015 de dispositivos médicos de Clase II, publicado el 8 de julio de 2015 e iniciado el 29 de mayo de 2015. Empresa responsable: Tecan US, Inc.; Estado: Finalizado; Cantidad reportada: 868 in total; Lote: Material Number 30019337. Serial numbers 1212004416 1009002605 612000003 811002510 712004158 906000965 902005109 912003472 1103002583 1411007090 704000003 702000001 708000500 1003004146 1109000103 1412005124 805001219 911005026 708005532 705007195 810000248 1010001663 712004152 708005533 1412005709 1103003780 706004455 608000002 908000185 1009002604 1308001035 804005914 1311007600 804000984 705009034 1309003195 801008055 706007961 901006341. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
A firmware bug causes incorrect values to be displayed for specific measured luminescence values.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Tecan US, Inc.
- Distribuidor
- US Nationwide Distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NE NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, including DC
- Lote
- Material Number 30019337. Serial numbers 1212004416 1009002605 612000003 811002510 712004158 906000965 902005109 912003472 1103002583 1411007090 704000003 702000001 708000500 1003004146 1109000103 1412005124 805001219 911005026 708005532 705007195 810000248 1010001663 712004152 708005533 1412005709 1103003780 706004455 608000002 908000185 1009002604 1308001035 804005914 1311007600 804000984 705009034 1309003195 801008055 706007961 901006341
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAGE Healthcare Precision 500D and Advantx Legacy/Legacy-D Radiographic and Fluoroscopic Systems. A universal8 de julio de 2015
- FDAGE Healthcare Discovery IGS 730. Angiographic X-ray systems used in generating fluoroscopic and rotational8 de julio de 2015
- FDAJustRight 5mm Reload, JR-REL25-2.0 Reload is packaged in cases of 12., placed in a tray and sealed with a8 de julio de 2015
- FDAGE Healthcare Optima CL 323i. Angiographic X-ray systems used in generating fluoroscopic and rotational images8 de julio de 2015
- FDAFreedom EVO-2 100 Base Unit, with Air LiHa Arms The Freedom EVO is an open automation platform product for5 de noviembre de 2014
- FDATecan Freedom EVO with Touch Tools Suite Version 3.0 The intended use of the Touch Tools Suite is to run26 de junio de 2013
Preguntas sobre este retiro
¿Infinite F500, in vitro diagnostic. Product Usage: The Tecan Infinite F500 is a multifunctional microplate reader with injector option. The Infinite F500 provides high performance for the vast majority of todays microplate applications and research and is robotic compatible. The Infinite F500 is intended as a general purpose laboratory instrument for professional use, supporting common microplate conforming to the ANS/SBS standards. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1967-2015 el 8 de julio de 2015. Peligro reportado: A firmware bug causes incorrect values to be displayed for specific measured luminescence values.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter