Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1971-2025

BD Texium Needle-Free Syringe: 3 mL, REF: MY8003; 5 mL, REF: MY8005; 10 mL, REF: MY8010; 20 mL, REF: MY8020; 30 mL, REF: MY8030; 60 mL, REF: MY8060

Publicado el 18 de junio de 2025 · Inicio del retiro: 12 de mayo de 2025

Resumen

FDA registra el retiro Z-1971-2025 de dispositivos médicos de Clase II, publicado el 18 de junio de 2025 e iniciado el 12 de mayo de 2025. Empresa responsable: BD SWITZERLAND SARL; Estado: En curso; Cantidad reportada: 1,265,400; Lote: UDI-DI/Lot(Expiration): MY8003/10885403221651/23125144(December 5, 2026), 24015468(January 20, 2027), 24025611(February 5, 2027), 24026390(February 27, 2027), 24036107(March 25, 2027), 24045647(April 24, 2027), 24055859(May 27, 2027), 24095513(September 24, 2027), 92217803(June 27, 2025), 92225801(September 15, 2025), 92231303(November 9, 2025), 92300201(January 2, 2026), 92308702(March 28, 2026), 92309501(April 5, 2026), 92313601(May 16, 2026), 92315701(June 6, 2026) MY8005/10885403221668/23125145(December 5, 2026), 24015467(January 20, 2027), 24026387(February 27, 2027), 24045646(April 24, 2027), 24055858(May 27, 2027), 24075164(July 26, 2027), 24095512(September 24, 2027), 92215201(June 1, 2025), 92220802(July 27, 2025), 92225502(September 12, 2025), 92315102(May 31, 2026); MY8010/10885403221675/24015312(January 15, 2027), 24025610(February 5, 2027), 24045645(April 24, 2027), 24055857(May 27, 2027), 24066953(June 25, 2027), 24075165(July 26, 2027), 24085486(August 26, 2027), 24095511(September 24, 2027), 24105167(October 25, 2027), 92218601(July 5, 2025), 92222902(August 17, 2025), 92223601(August 24, 2025), 92229201(October 19, 2025), 92230701(November 3, 2025), 92231301(November 9, 2025), 92231302(November 9, 2025), 92307402(March 15, 2026), 92311601(April 26, 2026), 92315101(May 31, 2026); MY8020/10885403221682/23115069(November 3, 2026), 24015311(January 15, 2027), 24025935(February 7, 2027), 24035566(March 5, 2027), 24036106(March 25, 2027), 24045644(April 24, 2027), 24055856(May 27, 2027), 24066952(June 25, 2027), 24075163(July 26, 2027), 24085485(August 26, 2027), 24095083(September 9, 2027), 24105196(October 30, 2027), 92216401(June 13, 2025), 92217801(June 27, 2025), 92218001(June 29, 2025), 92219202(July 11, 2025), 92231901(November 15, 2025), 92300205(January 2, 2026), 92300902(January 9, 2026), 92309503(April 5, 2026), 92310101(April 11, 2026), 92313002(May 10, 2026), 92313602(May 16, 2026) MY8030/10885403230523/23115070(November 3, 2026), 23125143(December 5, 2026), 24015264(January 12, 2027), 24015644(January 27, 2027), 24026389(February 27, 2027), 24036105(March 25, 2027), 24045643(April 24, 2027), 24055855(May 27, 2027), 24075162(July 26, 2027), 24085484(August 26, 2027), 24095509(September 24, 2027), 24115110(November 5, 2027), 92225501(September 12, 2025), 92233602(December 2, 2025), 92307403(March 15, 2026), 92308701(March 28, 2026), 92309502(April 5, 2026), 92315702(June 6, 2026), 92316402(June 13, 2026), 24066951(June 25, 2027); MY8060/10885403221699/23115496(November 28, 2026), 23125393(December 18, 2026), 24015102(January 8, 2027), 24026388(February 27, 2027), 24036104(March 25, 2027), 24045366(April 23, 2027), 24055854(May 27, 2027), 24075166(July 26, 2027), 24095596(September 24, 2027), 24115111(November 14, 2027), 92216402(June 13, 2025), 92222002(August 8, 2025), 92222901(August 17, 2025), 92224303(August 31, 2025), 92224301(August 31, 2025), 92226502(September 22, 2025), 92228501(October 12, 2025), 92230702(November 3, 2025), 92230703(November 3, 2025), 92230704(November 3, 2025), 92300202(January 2, 2026), 92300203(January 2, 2026), 92306703(March 8, 2026), 92306701(March 8, 2026), 92306702(March 8, 2026), 92311602(April 26, 2026), 92312302(May 3, 2026), 92313001(May 10, 2026), 92316601(June 15, 2026), 92316602(June 15, 2026). La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
BD SWITZERLAND SARL
Distribuidor
Worldwide - US Nationwide distribution including in the states of MA, CA, IL, AK, AL, MO, FL, VA, ME, NC, MD, IA, WI, AZ, CO, KY, IN, TX, AR, NJ, NY, MT, NE, TN, PA, GA, WA, KS, MN, ID, SC, MI, SD, OH, WV, LA, ND, VT, GU, NV, UT, OK, OR, DE, NM, MS, WY and the countries of CA, AU, NZ, CN.
Lote
UDI-DI/Lot(Expiration): MY8003/10885403221651/23125144(December 5, 2026), 24015468(January 20, 2027), 24025611(February 5, 2027), 24026390(February 27, 2027), 24036107(March 25, 2027), 24045647(April 24, 2027), 24055859(May 27, 2027), 24095513(September 24, 2027), 92217803(June 27, 2025), 92225801(September 15, 2025), 92231303(November 9, 2025), 92300201(January 2, 2026), 92308702(March 28, 2026), 92309501(April 5, 2026), 92313601(May 16, 2026), 92315701(June 6, 2026) MY8005/10885403221668/23125145(December 5, 2026), 24015467(January 20, 2027), 24026387(February 27, 2027), 24045646(April 24, 2027), 24055858(May 27, 2027), 24075164(July 26, 2027), 24095512(September 24, 2027), 92215201(June 1, 2025), 92220802(July 27, 2025), 92225502(September 12, 2025), 92315102(May 31, 2026); MY8010/10885403221675/24015312(January 15, 2027), 24025610(February 5, 2027), 24045645(April 24, 2027), 24055857(May 27, 2027), 24066953(June 25, 2027), 24075165(July 26, 2027), 24085486(August 26, 2027), 24095511(September 24, 2027), 24105167(October 25, 2027), 92218601(July 5, 2025), 92222902(August 17, 2025), 92223601(August 24, 2025), 92229201(October 19, 2025), 92230701(November 3, 2025), 92231301(November 9, 2025), 92231302(November 9, 2025), 92307402(March 15, 2026), 92311601(April 26, 2026), 92315101(May 31, 2026); MY8020/10885403221682/23115069(November 3, 2026), 24015311(January 15, 2027), 24025935(February 7, 2027), 24035566(March 5, 2027), 24036106(March 25, 2027), 24045644(April 24, 2027), 24055856(May 27, 2027), 24066952(June 25, 2027), 24075163(July 26, 2027), 24085485(August 26, 2027), 24095083(September 9, 2027), 24105196(October 30, 2027), 92216401(June 13, 2025), 92217801(June 27, 2025), 92218001(June 29, 2025), 92219202(July 11, 2025), 92231901(November 15, 2025), 92300205(January 2, 2026), 92300902(January 9, 2026), 92309503(April 5, 2026), 92310101(April 11, 2026), 92313002(May 10, 2026), 92313602(May 16, 2026) MY8030/10885403230523/23115070(November 3, 2026), 23125143(December 5, 2026), 24015264(January 12, 2027), 24015644(January 27, 2027), 24026389(February 27, 2027), 24036105(March 25, 2027), 24045643(April 24, 2027), 24055855(May 27, 2027), 24075162(July 26, 2027), 24085484(August 26, 2027), 24095509(September 24, 2027), 24115110(November 5, 2027), 92225501(September 12, 2025), 92233602(December 2, 2025), 92307403(March 15, 2026), 92308701(March 28, 2026), 92309502(April 5, 2026), 92315702(June 6, 2026), 92316402(June 13, 2026), 24066951(June 25, 2027); MY8060/10885403221699/23115496(November 28, 2026), 23125393(December 18, 2026), 24015102(January 8, 2027), 24026388(February 27, 2027), 24036104(March 25, 2027), 24045366(April 23, 2027), 24055854(May 27, 2027), 24075166(July 26, 2027), 24095596(September 24, 2027), 24115111(November 14, 2027), 92216402(June 13, 2025), 92222002(August 8, 2025), 92222901(August 17, 2025), 92224303(August 31, 2025), 92224301(August 31, 2025), 92226502(September 22, 2025), 92228501(October 12, 2025), 92230702(November 3, 2025), 92230703(November 3, 2025), 92230704(November 3, 2025), 92300202(January 2, 2026), 92300203(January 2, 2026), 92306703(March 8, 2026), 92306701(March 8, 2026), 92306702(March 8, 2026), 92311602(April 26, 2026), 92312302(May 3, 2026), 92313001(May 10, 2026), 92316601(June 15, 2026), 92316602(June 15, 2026)
Clasificación de riesgo
Class II
Área (FDA)
device
País
Switzerland

Retiros relacionados

Preguntas sobre este retiro

¿BD Texium Needle-Free Syringe: 3 mL, REF: MY8003; 5 mL, REF: MY8005; 10 mL, REF: MY8010; 20 mL, REF: MY8020; 30 mL, REF: MY8030; 60 mL, REF: MY8060 fue retirado del mercado?

Sí. FDA publicó el retiro Z-1971-2025 el 18 de junio de 2025. Peligro reportado: Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter