Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1973-2025

BD TEXIUM CLOSED MALE LUER WITH FEMALE CAP, REF: 10012241-0500

Publicado el 18 de junio de 2025 · Inicio del retiro: 12 de mayo de 2025

Resumen

FDA registra el retiro Z-1973-2025 de dispositivos médicos de Clase II, publicado el 18 de junio de 2025 e iniciado el 12 de mayo de 2025. Empresa responsable: BD SWITZERLAND SARL; Estado: En curso; Cantidad reportada: 1,655,300; Lote: REF/UDI-DI/Lot(Expiration): 10012241-0500/10885403221705/23045145(April 3, 2026), 23045146(April 3, 2026), 23055091(May 4, 2026), 23065090(June 3, 2026), 23075107(July 5, 2026), 23095124(September 8, 2026), 23105027(October 3, 2026), 23115047(November 2, 2026), 23125146(December 5, 2026), 24015245(January 11, 2027), 24025887(February 7, 2027), 24026421(February 27, 2027), 24036126(March 25, 2027), 24045690(April 24, 2027), 24066893(June 25, 2027), 24075111(July 26, 2027), 24095556(September 24, 2027), 24125014(December 2, 2027), 23035049(March 1, 2026). La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
BD SWITZERLAND SARL
Distribuidor
Worldwide - US Nationwide distribution including in the states of MA, CA, IL, AK, AL, MO, FL, VA, ME, NC, MD, IA, WI, AZ, CO, KY, IN, TX, AR, NJ, NY, MT, NE, TN, PA, GA, WA, KS, MN, ID, SC, MI, SD, OH, WV, LA, ND, VT, GU, NV, UT, OK, OR, DE, NM, MS, WY and the countries of CA, AU, NZ, CN.
Lote
REF/UDI-DI/Lot(Expiration): 10012241-0500/10885403221705/23045145(April 3, 2026), 23045146(April 3, 2026), 23055091(May 4, 2026), 23065090(June 3, 2026), 23075107(July 5, 2026), 23095124(September 8, 2026), 23105027(October 3, 2026), 23115047(November 2, 2026), 23125146(December 5, 2026), 24015245(January 11, 2027), 24025887(February 7, 2027), 24026421(February 27, 2027), 24036126(March 25, 2027), 24045690(April 24, 2027), 24066893(June 25, 2027), 24075111(July 26, 2027), 24095556(September 24, 2027), 24125014(December 2, 2027), 23035049(March 1, 2026)
Clasificación de riesgo
Class II
Área (FDA)
device
País
Switzerland

Retiros relacionados

Preguntas sobre este retiro

¿BD TEXIUM CLOSED MALE LUER WITH FEMALE CAP, REF: 10012241-0500 fue retirado del mercado?

Sí. FDA publicó el retiro Z-1973-2025 el 18 de junio de 2025. Peligro reportado: Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter