FDAFDA · Alimentos y medicamentos·N.º oficial Z-1981-2012
Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-2 (straight handle) Stainless Steel / Silicone Rubber, Rx only, non sterile. Product Usage: PathFinder NXT is a minimally invasive pedicle screw system. The PathFinder NXT System consists of various screws and rods intended to provide temporary stabilization following surgery to fuse the spine. The intended use of the Fixed Angled Percutaneous Rod Inserter is to insert a rod percutaneously through the extender sieeves to position into the implanted screw heads for securing with closure tops forming a construct that provides temporary stabilization following MIS or mini-open surgery to fuse the spine.
Publicado el 18 de julio de 2012 · Inicio del retiro: 5 de junio de 2012
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Zimmer is recalling Pathfinder NXT Fixed Percutaneous Rod Holder, because complaints have been received that the rod holder tip may crack or break when the rod is manipulated during surgery.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Zimmer, Inc.
- Distribuidor
- Worldwide Distribution - USA (nationwide) and the countries of: Austria, Australia, Belgium, France, Gelmany, Italy, Spain, Switzerland and the United Kingdom.
- Lote
- Part numbers 3573-2 (straight handle) all lots.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States