Synthes 2.4mm VA-LCP Two Column Volar Distal Radius Plate and Torque limiting attachment, 1.2Nm Synthes VA-LCP Distal Radius System is intended for fixation of complex intra- and extra-articular fractures and osteotomies of the distal radius and other small bones.
Publicado el 4 de septiembre de 2013 · Inicio del retiro: 31 de mayo de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The firm is initiating a Field Safety Correction for the labeling of the 2.4mm VA-LCP Two Column Volar Distal Radius Plate and 1.2 Nm Torque Limiting Attachment. The US Technique guide lists 1.2Nm TLA as an option for use with the 2.4mm Variable Angle LCP (VA-LCP) Two-Column (Standard & Narrow) Volar Distal Radius Plate. The potential exists for the VA locking screw to be driven through the 2.4mm
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Synthes (USA) Products LLC
- Distribuidor
- Nationwide Distribution including the states of AK. AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MD, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI SC, SD, TN, TX, UT, VA, VT, WA WI and WV.
- Lote
- Part number J8301-F
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States