Affinity Four Birthing Bed P3700 Product Usage: The Affinity, Four Birthing Bed is intended to be used as birthing beds for women of child bearing age in an LDR (Labor, Delivery, Recovery ) or LDRP (Labor, Delivery, Recovery, Postpartum) setting within the acute care Labor and Delivery market. They are not intended for use as general hospital beds.
Publicado el 1 de agosto de 2012 · Inicio del retiro: 2 de abril de 2012
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
During review of our last braking system via that field corrective action, the system still did not meet our expectations, which is to eliminate the hazards of patient falling. Bed movement immediately prior to, or during the birthing process may still create distractions and delays, resulting in patient or caregiver falls, or contribute to a deviation in a procedure, (e.g. IV insertion, epidural
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Hill-Rom, Inc.
- Distribuidor
- Worldwide Distribution - US Nationwide and countries including: Argentina , Aruba, Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Brazil, Brunei, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Finland, France, Germany, Greece, Guam, Guyana, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Lebanon, Libya, Malaysia, Mauritius, Mexico, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad and Tobago, Turkey, Turks and Caicos Islands, United Arab Emirates, United Kingdom, United States, US Virgin Islands, Venezuela and Vietnam
- Lote
- Affinity Four Birthing Beds (P3700) manufactured with Serial Numbers H331AA4785 thru N081AA7966
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States