Extraneal and/or DIANEAL LOW CALCIUM Peritoneal Dialysis Solution in Plastic Container, PL-146 used with Automated Peritoneal Dialysis System/Automated PD Set with Cassette; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A. The HomeChoice dialysis system is a personal cycler automated peritoneal dialysis system for pediatric and adult renal patients with fill volumes ranging from 60 mL to 3000 mL. DIANEAL Low Calcium peritoneal dialysis solutions are indicated for use in chronic renal failure patients being maintained on peritoneal dialysis. - and- The Personal Cycler Peritoneal Dialysis System [cycler and ancillary sets] is used for automatic control of dialysate solution exchanges in treatment of renal failure patients undergoing peritoneal dialysis. The Personal Cycler System automatically cycles a prescribed volume of dialysis solution into and out of the peritoneal cavity during the dialysis treatment.
Publicado el 8 de agosto de 2012 · Inicio del retiro: 6 de julio de 2012
Resumen
FDA registra el retiro Z-2132-2012 de dispositivos médicos de Clase II, publicado el 8 de agosto de 2012 e iniciado el 6 de julio de 2012. Empresa responsable: Baxter Healthcare Renal Div; Estado: Finalizado; Cantidad reportada: N/A; Lote: all lots and serial numbers. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Baxter discovered that a Baxter technical service representative was providing incorrect instructions for inspection of solution bags for a blocked frangible prior to connection to the Automated PD Set with Cassette.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Baxter Healthcare Renal Div
- Distribuidor
- Nationwide Distribution includingArizona, California, Ohio, Pennsylvania and Virginia.
- Lote
- all lots and serial numbers
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAFlush Kit w 3 Way Off Stopcock Product Usage: Flush Kit8 de agosto de 2012
- FDAI-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ. The i-Stat EG7 cartridge is intended to be used for8 de agosto de 2012
- FDA94" (239cm) 20 Drop Admin Set w/High Pressure 4-Way (300 psi) Stopcock, Rotating Luer, Non-DEHP Tubing8 de agosto de 2012
- FDARF Cannula, 100 mm,Curved 5 MM, 20 G, REF 0406-630-115 RX only, Disposable ( 10/pk) Product Usage: The Stryker8 de agosto de 2012
- FDACardiac Catheterization Waste Management System w/1000mL Waste Bag, 2 Fluid Admin Sets Product Usage: For8 de agosto de 2012
- FDARF Cannula, 100 mm,Curved 10 MM, 20 G, REF 0406-630-125 RX only, Disposable ( 10/pk) Product Usage: The8 de agosto de 2012
Preguntas sobre este retiro
¿Extraneal and/or DIANEAL LOW CALCIUM Peritoneal Dialysis Solution in Plastic Container, PL-146 used with Automated Peritoneal Dialysis System/Automated PD Set with Cassette; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A. The HomeChoice dialysis system is a personal cycler automated peritoneal dialysis system for pediatric and adult renal patients with fill volumes ranging from 60 mL to 3000 mL. DIANEAL Low Calcium peritoneal dialysis solutions are indicated for use in chronic renal failure patients being maintained on peritoneal dialysis. - and- The Personal Cycler Peritoneal Dialysis System [cycler and ancillary sets] is used for automatic control of dialysate solution exchanges in treatment of renal failure patients undergoing peritoneal dialysis. The Personal Cycler System automatically cycles a prescribed volume of dialysis solution into and out of the peritoneal cavity during the dialysis treatment. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2132-2012 el 8 de agosto de 2012. Peligro reportado: Baxter discovered that a Baxter technical service representative was providing incorrect instructions for inspection of solution bags for a blocked frangible prior to connection to the Automated PD Set with Cassette.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter