Kodak DirectView DR 7500 Diagnostic X-Ray System Product Usage: The DR 7500 system is a permanently installed diagnostic x-ray system composed of 4 main components: an operator console, over head tube crane, x-ray tube assembly including a collimator and stationary generator. In addition the device can use an x-ray table and /or wallstand/bucky to complete x-ray exposures. The DR 7500 System is a permanently installed diagnostic x-ray system for generation of x-rays for examination of various anatomical regions.
Publicado el 20 de agosto de 2014 · Inicio del retiro: 9 de julio de 2014
Resumen
FDA registra el retiro Z-2177-2014 de dispositivos médicos de Clase II, publicado el 20 de agosto de 2014 e iniciado el 9 de julio de 2014. Empresa responsable: Carestream Health, Inc.; Estado: Finalizado; Cantidad reportada: 589 systems; Lote: The device can be tracked according to service code. The service code for the DR 7500 system is 8087. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
An on-site evaluation of the equipment parts that attach the Beta Assembly (x-ray tube and collimator) to the Over Head Tube Crane (OTC) Telescope was completed by Carestream Health (CSH) Service Engineering. Inspection of these parts revealed that the tab on the lock washer was not bent into the lock nut recess and because the Beta Assembly detached from the OTC telescope, the lock nut was not sufficiently tightened.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Carestream Health, Inc.
- Distribuidor
- Worldwide Distribution - US Nationwide and the countries of (United Kingdom, United Arab Emirates, Turkey, Taiwan, Switzerland, Sweden, Spain, South Africa, Portugal, Poland, Oman, New Zealand, Netherlands, Kuwait, Italy, Israel, Hong Kong, Germany, France, Finland, Denmark, Dubai, Czech Republic, China, Korea, Belgium, Australia, and Amsterdam.
- Lote
- The device can be tracked according to service code. The service code for the DR 7500 system is 8087.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDADRX Revolution Mobile X-Ray System13 de diciembre de 2023
- FDAOnSight 3D Extremity System- X-Ray, Tomography Computed System Catalog Numbers: 1743566 (w/ Attached Console)1 de mayo de 2019
- FDACARESTREAM DIRECTVIEW CR Software Generates digital mammographic images that can be used for screening and24 de diciembre de 2014
- FDAAgfa Digital Radiography X-Ray System DX-D 100 Product Usage: Is indicated for use in providing diagnostic20 de agosto de 2014
- FDARadiomat M+ NIF 14 x 17; Radiomat SG 14 x 17 NIF; RX-B 8 x 10 in./IBF-004355; Radiomat B+ NIF 14 x 17;20 de agosto de 2014
- FDAKODAK DirectView DR 7500 Dual Detector System, MODEL DR 7500, Manufactured by Carestream Health, Inc. Made in5 de febrero de 2014
Preguntas sobre este retiro
¿Kodak DirectView DR 7500 Diagnostic X-Ray System Product Usage: The DR 7500 system is a permanently installed diagnostic x-ray system composed of 4 main components: an operator console, over head tube crane, x-ray tube assembly including a collimator and stationary generator. In addition the device can use an x-ray table and /or wallstand/bucky to complete x-ray exposures. The DR 7500 System is a permanently installed diagnostic x-ray system for generation of x-rays for examination of various anatomical regions. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2177-2014 el 20 de agosto de 2014. Peligro reportado: An on-site evaluation of the equipment parts that attach the Beta Assembly (x-ray tube and collimator) to the Over Head Tube Crane (OTC) Telescope was completed by Carestream Health (CSH) Service Engineering. Inspection of these parts revealed that the tab on the lock washer was not bent into the lock nut recess and because the Beta Assembly detached from the OTC telescope, the lock nut was not sufficiently tightened.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter