Exact Sciences Sample Mixer 2. For In Vitro Diagnostic Use. The Exact Sciences Sample Mixer 2 is intended for mixing stool samples in a Sample Container as part of pre-analytic sample processing for in vitro diagnostic tests

Publicado el 7 de junio de 2017 · Inicio del retiro: 28 de abril de 2017

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Exact Sciences is correcting all Sample Mixer v2 Upgrade units because of incidents of Sample Mixer door damage, which has occurred when the Sample Holder opens during a run and the Sample Holder lid makes contact with the Sample Mixer door.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Exact Sciences Corporation
Distribuidor
US to WI only
Lote
Equipment ID / Serial Number: MIX-0051 / MX00002051, MIX-0052 / MX00002032, MIX-0053 / MX00002028, MIX-0054 / MX00002052, MIX-0055 / MX00002043, MIX-0056 / MX00002040, MIX-0057 / MX00002050, MIX-0058 / MX00002036, MIX-0059 / MX00002027, MIX-0060 / MX00002015, MIX-0061 / MX00002016, MIX-0063 / MX00002039, MIX-0064 / MX00002033, MIX-0065 / MX00002053, MIX-0066 / MX00002014,