Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2232-2014

VIDAS D-Dimer Exclusion II, an automated quantitative test for use on instruments of the VIDAS family for the immunoenzymatic determination of fibrin degradation products in human plasma using the Enzyme Linked Fluorescent Assay technique. Also indicated for use in conjunction with clinical pretest probability assessment model to exclude DVT and PE disease.

Publicado el 20 de agosto de 2014 · Inicio del retiro: 17 de abril de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Customer complaints about not repeatable results below the detection limit (<45 ng/mL). Therefore false negatives are being reported for results around the medical decision point of 500 ng/mL that initially tested at <45 ng/mL.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
BioMerieux SA
Distribuidor
AL, AR, CA, CO, CT, DE, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MA, MI, MN, MO, MT, NE, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, VT, VA, WA, WV, WI, WY
Lote
Reference 304551-01 - Lots 1001625210, Exp. date 8/8/20123; 1001715580, Exp. date 9/26/2013; 1001711530, Exp. date 9/11/2013; 1001789540, Exp. date 10/29/2013; 1001932680, Exp. date 12/12/2013
Clasificación de riesgo
Class II
Área (FDA)
device
País
France

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta