VIDAS D-Dimer Exclusion II, an automated quantitative test for use on instruments of the VIDAS family for the immunoenzymatic determination of fibrin degradation products in human plasma using the Enzyme Linked Fluorescent Assay technique. Also indicated for use in conjunction with clinical pretest probability assessment model to exclude DVT and PE disease.
Publicado el 20 de agosto de 2014 · Inicio del retiro: 17 de abril de 2013
Resumen
FDA registra el retiro Z-2232-2014 de dispositivos médicos de Clase II, publicado el 20 de agosto de 2014 e iniciado el 17 de abril de 2013. Empresa responsable: BioMerieux SA; Estado: Finalizado; Cantidad reportada: 2160 units; Lote: Reference 304551-01 - Lots 1001625210, Exp. date 8/8/20123; 1001715580, Exp. date 9/26/2013; 1001711530, Exp. date 9/11/2013; 1001789540, Exp. date 10/29/2013; 1001932680, Exp. date 12/12/2013. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Customer complaints about not repeatable results below the detection limit (<45 ng/mL). Therefore false negatives are being reported for results around the medical decision point of 500 ng/mL that initially tested at <45 ng/mL.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- BioMerieux SA
- Distribuidor
- AL, AR, CA, CO, CT, DE, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MA, MI, MN, MO, MT, NE, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, VT, VA, WA, WV, WI, WY
- Lote
- Reference 304551-01 - Lots 1001625210, Exp. date 8/8/20123; 1001715580, Exp. date 9/26/2013; 1001711530, Exp. date 9/11/2013; 1001789540, Exp. date 10/29/2013; 1001932680, Exp. date 12/12/2013
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- France
Retiros relacionados
- FDAEMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation4 de noviembre de 2026
- FDAMYLA software. Used to manage microbiology test workflow from the reception of requests to the transmission of18 de agosto de 2021
- FDAVIDAS FSH, Ref 30407-01, IVD, Rx. The firm name on the label is bioMerieux SA, France. For use on the4 de julio de 2018
- FDAThe VITEK® 2 Gram Positive Susceptibility Card is intended for use with the VITEK 2 Systems in clinical20 de junio de 2018
- FDA23.5 Deg Rasp Handle Assy- RIGHT Nonsterile20 de agosto de 2014
- FDABD Multi Check CD4 Low Control Catalog No. 340916 is intended as a complete process control for20 de agosto de 2014
Preguntas sobre este retiro
¿VIDAS D-Dimer Exclusion II, an automated quantitative test for use on instruments of the VIDAS family for the immunoenzymatic determination of fibrin degradation products in human plasma using the Enzyme Linked Fluorescent Assay technique. Also indicated for use in conjunction with clinical pretest probability assessment model to exclude DVT and PE disease. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2232-2014 el 20 de agosto de 2014. Peligro reportado: Customer complaints about not repeatable results below the detection limit (<45 ng/mL). Therefore false negatives are being reported for results around the medical decision point of 500 ng/mL that initially tested at <45 ng/mL.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter