GE Healthcare Discovery IGS 730 Cardiovascular X-Ray Imaging System. The angiography X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures. Additionally, with the OR table, the angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for image guided surgical procedures. The OR table is suitable for interventional and surgical procedures.
Publicado el 25 de septiembre de 2013 · Inicio del retiro: 19 de junio de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
GE Healthcare has recently become aware of a potential safety issue with respect to Discovery IGS 730 Imaging Systems. A patient data management error may prevent the Discovery IGS 730 Imaging System from booting on start-up or re-booting during procedure. The boot issue may occur when a large amount of patient data remains in the browser and is not deleted. The inability to boot the system ma
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- GE Healthcare, LLC
- Distribuidor
- Worldwide distribution: US (nationwide) including states of : PA and TN; countries of: FRANCE and SINGAPORE.
- Lote
- Mfg Lot or Serial # 00000611882BU2 00000612803BU7 00000613979BU4 00000620447BU3
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States