Whole Body X--ray Scanner This device is indicated as a whole body volume CT system. The system produces volumes of up to 320 axial slices in a single rotation.

Publicado el 5 de septiembre de 2012 · Inicio del retiro: 13 de julio de 2009

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

1. It has been found that in raw data processing for Go&Return helical scanning, when the raw data acquired in the first scan (Go) and that acquired in the second scan (Return) are selected and reconstructed at one time, incorrect couch positions and scan times are displayed on the reconstructed images for the data acquired by Return scanning, while all related information is displayed on the reco

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Toshiba American Medical Systems Inc
Distribuidor
Nationwide distribution: USA including states of: AL, AR, CA, CO, FL, GA, KY, MA, MD, MT, ND, NJ, NV, NY, OH, PA and Puerto Rico.
Lote
Serial Number: 2DA0882051
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States