The FL23SE Bed is an AC-powered adjustable hospital bed with four built-in electric DC motors and remote controls that is intended for medical purposes. It can be operated by the patient or the caregiver to adjust the height and surface contour of the bed. The bed includes movable and latchable siderails. The FL23SE Bed is intended for use with patients for procedures, therapy, recovery in healthcare environments, and to transport patients between bays and procedural rooms. The FL23SE Bed is intended for use with patients for procedures, therapy, recovery in healthcare environments, and to transport patients between bays and procedural rooms.
Publicado el 2 de octubre de 2013 · Inicio del retiro: 19 de agosto de 2013
Resumen
FDA registra el retiro Z-2283-2013 de dispositivos médicos de Clase II, publicado el 2 de octubre de 2013 e iniciado el 19 de agosto de 2013. Empresa responsable: Stryker Medical Division of Stryker Corporation; Estado: Finalizado; Cantidad reportada: 61 units; Lote: Model Numbers: FL23SE with nurse call option: OL235017 & OL235018 List of Affected Devices: J60267 J60425 J60009 J60001 J60094 J60095 J60096 J60348 J60349 J60350 J60351 J60352 J60353 J60354 J60360 J60361 J60362 J60363 J60364 J60365 J60366 J60367 J60368 J60369 J60370 J60371 J60372 J60373 J60374 J60375 J60376 J60377 J60378 J60379 J60380 J60381 J60382 J60383 J60384 J60385 J60386 J60387 J60388 J60389 J60390 J60391 J60392 J60393 J60394 J60395 J60396 J60397 J60398 J60399 J60400 J60401 J60402 J60403 J60404 J60405 J60512. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The nurse call cable (Part Number: QDF23-0571) on the FL23SE Bed is located right under the opening of the bed head box. If an accessory (i.e. IV pole) is inserted into this slot, the circuit directing the nurse call system can be damaged.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Stryker Medical Division of Stryker Corporation
- Distribuidor
- Worldwide Distribution - USA including MI, WI, VA and internationally to Canada.
- Lote
- Model Numbers: FL23SE with nurse call option: OL235017 & OL235018 List of Affected Devices: J60267 J60425 J60009 J60001 J60094 J60095 J60096 J60348 J60349 J60350 J60351 J60352 J60353 J60354 J60360 J60361 J60362 J60363 J60364 J60365 J60366 J60367 J60368 J60369 J60370 J60371 J60372 J60373 J60374 J60375 J60376 J60377 J60378 J60379 J60380 J60381 J60382 J60383 J60384 J60385 J60386 J60387 J60388 J60389 J60390 J60391 J60392 J60393 J60394 J60395 J60396 J60397 J60398 J60399 J60400 J60401 J60402 J60403 J60404 J60405 J60512
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAProCuity bed series, model number 3009, item number: 3009PX-LE-400.13 de diciembre de 2023
- FDAProform- support surface is used to assist in the prevention and treatment of all categories/stages of18 de agosto de 2021
- FDAPrime Fifth Wheel Stretcher-stretcher may be used for minor procedures and short-term stay (treatment and18 de agosto de 2021
- FDAThe Spirit Plus Bed, Model 5600 Spirit Plus /with high rail siderail option21 de noviembre de 2018
- FDASynthes Oracle Spacer System Slap Hammer, Part Number 03.809.690, Lot Number 3723847 The Oracle Spacer is2 de octubre de 2013
- FDAUniversal Titanium Prosthesis, Round Head, with Flex H/A titanium shoe. The ear prosthesis is implantable and2 de octubre de 2013
Preguntas sobre este retiro
¿The FL23SE Bed is an AC-powered adjustable hospital bed with four built-in electric DC motors and remote controls that is intended for medical purposes. It can be operated by the patient or the caregiver to adjust the height and surface contour of the bed. The bed includes movable and latchable siderails. The FL23SE Bed is intended for use with patients for procedures, therapy, recovery in healthcare environments, and to transport patients between bays and procedural rooms. The FL23SE Bed is intended for use with patients for procedures, therapy, recovery in healthcare environments, and to transport patients between bays and procedural rooms. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2283-2013 el 2 de octubre de 2013. Peligro reportado: The nurse call cable (Part Number: QDF23-0571) on the FL23SE Bed is located right under the opening of the bed head box. If an accessory (i.e. IV pole) is inserted into this slot, the circuit directing the nurse call system can be damaged.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter