Merlin PCS programmer (Model # 3650), Software (Model # 3330)
Publicado el 4 de julio de 2018 · Inicio del retiro: 16 de abril de 2018
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against unauthorized device access. For older generation devices not capable of accepting the firmware, the firm is advising customers of the option to permanently disable the RF communication capability in these devices.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- St Jude Medical Inc.
- Distribuidor
- U.S. Nationwide distribution
- Lote
- Software Versions: 24.1.1 rev1 (US), 23.1.2 rev 1 (OUS), 23.1.3 (Japan), and 23.1.4 rev 1 (Canada)
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States