FDAFDA · Alimentos y medicamentos·N.º oficial Z-2360-2016

Alere INRatio2 PT/INR Professional Monitoring System Model Number: 55128A Product Usage: The Alere INRatio¿/INRatio¿2 system uses a modified version of the one-stage PT test. After a drop of blood is applied to the test strip, it is drawn into the test area and mixed with reagents that initiate coagulation. When the blood clots, changes in impedance in the sample are detected by the monitor. The monitor calculates the PT and INR results from this impedance change and reports them on the display. The Alere INRatio/INRatio2 PT/INR Monitoring System is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The Alere INRatio/INRatio2 PT/INR Monitoring System is intended for use outside the body (in vitro diagnostic use). The Alere INRatio/INRatio2 PT/INR Monitoring System is intended for professional and home use by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The Alere INRatio/INRatio2 PT/INR Monitoring System is not intended to be used for screening purposes.

Publicado el 31 de agosto de 2016 · Inicio del retiro: 11 de julio de 2016

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Qué hacer

Voluntary: Firm initiated · Press Release

Detalles del retiro

Fabricante
Alere San Diego, Inc.
Distribuidor
Worldwide Distribution - US Nationwide and foreign distribution to the countries of: Argentina, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Brunei Daruss, Canada, Chile, China, Colombia, Egypt, France, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kazakhstan, Malaysia, Maldives, Netherlands, Russia, Rwanda, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, Trinidad, Ukraine, Uruguay, and Vietnam.
Lote
All Lots All Serial Numbers
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States