Alere INRatio PT/INR Test Strips Model Number: 0100071, 0100139 Product Usage: The INRatio/INRatio2 Test Strips perform a modified version of the one-stage Prothrombin Time test, using a recombinant human thromboplastin reagent. The clot formed in the Prothrombin Time reaction is detected by a change in the electrical impedance of the sample during the coagulation process. The system consists of a monitor and disposable test strips. The monitor measures impedance, heats the test strip to the proper reaction temperature, and provides a user interface. The blood sample is applied to the Test Strip and the clotting reaction occurs on the Test Strip.
Publicado el 31 de agosto de 2016 · Inicio del retiro: 11 de julio de 2016
Resumen
FDA registra el retiro Z-2361-2016 de dispositivos médicos de Clase I, publicado el 31 de agosto de 2016 e iniciado el 11 de julio de 2016. Empresa responsable: Alere San Diego, Inc.; Estado: Finalizado; Lote: All Lots All Serial Numbers. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Alere San Diego is initiating a voluntary recall for the Alere INRatio®/INRatio®2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio®/INRatio®2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
Qué hacer
Voluntary: Firm initiated · Press Release
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Alere San Diego, Inc.
- Distribuidor
- Worldwide Distribution - US Nationwide and foreign distribution to the countries of: Argentina, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Brunei Daruss, Canada, Chile, China, Colombia, Egypt, France, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kazakhstan, Malaysia, Maldives, Netherlands, Russia, Rwanda, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, Trinidad, Ukraine, Uruguay, and Vietnam.
- Lote
- All Lots All Serial Numbers
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAFisher Scientific Sure-Vue Urine hCG Strip, Model/ Part: FHC-101/23900527; McKesson CONSULT hCG Urine Test7 de octubre de 2026
- FDAMcKesson CONSULT hCG Urine Cassette, Model/ Part: FHC-A102/5001; Alere hCG Cassette (20 mIU/mL) in Urine,7 de octubre de 2026
- FDAFisher Scientific Sure-Vue Urine hCG, Model/ Part: FHC-102/23900526; Henry Schein, Inc. One Step + Urine7 de octubre de 2026
- FDAFisher Scientific Sure-Vue Serum/Urine hCG, Model/ Part: FHC-202/23900529, FHC-202/23900528; Beckman Coulter7 de octubre de 2026
- FDAAlere INRatio2 PT/INR Professional Monitoring System Model Number: 55128A Product Usage: The Alere31 de agosto de 2016
- FDAAdult ThermoGard® Dual Dispersive Electrodes (for patients >15 kg.), 10' (3.05m) Cable, Catalog Number31 de agosto de 2016
Preguntas sobre este retiro
¿Alere INRatio PT/INR Test Strips Model Number: 0100071, 0100139 Product Usage: The INRatio/INRatio2 Test Strips perform a modified version of the one-stage Prothrombin Time test, using a recombinant human thromboplastin reagent. The clot formed in the Prothrombin Time reaction is detected by a change in the electrical impedance of the sample during the coagulation process. The system consists of a monitor and disposable test strips. The monitor measures impedance, heats the test strip to the proper reaction temperature, and provides a user interface. The blood sample is applied to the Test Strip and the clotting reaction occurs on the Test Strip. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2361-2016 el 31 de agosto de 2016. Peligro reportado: Alere San Diego is initiating a voluntary recall for the Alere INRatio®/INRatio®2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio®/INRatio®2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Press Release