Ingenia 1.5T CX-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body, or extremities Models: 781261 781262
Publicado el 15 de septiembre de 2021 · Inicio del retiro: 11 de agosto de 2021
Resumen
FDA registra el retiro Z-2414-2021 de dispositivos médicos de Clase II, publicado el 15 de septiembre de 2021 e iniciado el 11 de agosto de 2021. Empresa responsable: Philips North America Llc; Estado: Completado; Cantidad reportada: 12 units, 137 units; Lote: Model: 781261 Serial Number Unique Device Identifier (UDI#) 77008 77000 77004 77007 77003 77001 77002 77018 77016 77019 77010 77014 Model:781262 77131 (01)00884838068445(21)77131 77031 N/A 77046 N/A 77070 N/A 77124 (01)00884838068445(21)77124 77144 (01)00884838068445(21)77144 77172 (01)00884838068445(21)77172 77074 (01)00884838068445(21)77074 77058 N/A 77114 N/A 77127 (01)00884838068445(21)77127 77160 (01)00884838068445(21)77160 77159 (01)00884838068445(21)77159 77182 (01)00884838068445(21)77182 77180 (01)00884838068445(21)77180 77171 (01)00884838068445(21)77171 77179 (01)00884838068445(21)77179 77048 N/A 77040 N/A 77034 N/A 77039 N/A 77035 N/A 77092 N/A 77140 (01)00884838068445(21)77140 77177 (01)00884838068445(21)77177 77181 (01)00884838068445(21)77181 77178 (01)00884838068445(21)77178 77041 N/A 77103 N/A 77193 (01)00884838068445(21)77193 77203 (01)00884838068445(21)77203 77134 (01)00884838068445(21)77134 77047 N/A 77051 N/A 77050 N/A 77060 N/A 77194 (01)00884838068445(21)77194 77111 N/A 77119 (01)00884838068445(21)77119 77167 (01)00884838068445(21)77167 77183 (01)00884838068445(21)77183 77185 (01)00884838068445(21)77185 77187 (01)00884838068445(21)77187 77188 (01)00884838068445(21)77188 77198 (01)00884838068445(21)77198 77195 (01)00884838068445(21)77195 77196 (01)00884838068445(21)77196 77199 N/A 77202 N/A 77037 N/A 77057 N/A 77072 N/A 77075 N/A 77084 N/A 77085 N/A 77078 N/A 77086 N/A 77101 N/A 77110 (01)00884838068445(21)77110 77038 (01)00884838068445(21)77038 77157 (01)00884838068445(21)77157 77156 (01)00884838068445(21)77156 77056 N/A 77052 N/A 77054 N/A 77053 N/A 77055 N/A 77064 N/A 77065 N/A 77067 N/A 77066 N/A 77068 N/A 77069 N/A 77105 N/A 77126 (01)00884838068445(21)77126 77125 (01)00884838068445(21)77125 77137 (01)00884838068445(21)77137 77139 (01)00884838068445(21)77139 77136 (01)00884838068445(21)77136 77135 (01)00884838068445(21)77135 77138 (01)00884838068445(21)77138 77148 (01)00884838068445(21)77148 77141 (01)00884838068445(21)77141 77030 N/A 77033 N/A 77036 N/A 77042 N/A 77044 N/A 77043 N/A 77032 N/A 77049 N/A 77061 N/A 77059 N/A 77071 N/A 77090 N/A 77088 N/A 77093 N/A 77091 N/A 77087 N/A 77097 N/A 77083 N/A 77106 N/A 77112 N/A 77107 (01)00884838068445(21)77107 77115 N/A 77100 (01)00884838068445(21)77100 77118 (01)00884838068445(21)77118 77121 (01)00884838068445(21)77121 77129 (01)00884838068445(21)77129 77120 (01)00884838068445(21)77120 77128 (01)00884838068445(21)77128 77151 (01)00884838068445(21)77151 77143 (01)00884838068445(21)77143 77158 (01)00884838068445(21)77158 77162 (01)00884838068445(21)77162 77161 (01)00884838068445(21)77161 77155 (01)00884838068445(21)77155 77190 (01)00884838068445(21)77190 77191 (01)00884838068445(21)77191 77150 (01)00884838068445(21)77150 77081 N/A 77117 (01)00884838068445(21)77117 77123 (01)00884838068445(21)77123 77166 (01)00884838068445(21)77166 77045 N/A 77063 N/A 77109 N/A 77201 (01)00884838068445(21)77201 77113 N/A 77147 (01)00884838068445(21)77147 77122 (01)00884838068445(21)77122 77154 (01)00884838068445(21)77154 77169 (01)00884838068445(21)77169 77168 (01)00884838068445(21)77168 77130 (01)00884838068445(21)77130 77094 N/A 77062 N/A. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Completed — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Philips North America Llc
- Distribuidor
- Worldwide distribution - US Nationwide distribution and the countries of Albania, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan Palestine, State of Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R®union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam.
- Lote
- Model: 781261 Serial Number Unique Device Identifier (UDI#) 77008 77000 77004 77007 77003 77001 77002 77018 77016 77019 77010 77014 Model:781262 77131 (01)00884838068445(21)77131 77031 N/A 77046 N/A 77070 N/A 77124 (01)00884838068445(21)77124 77144 (01)00884838068445(21)77144 77172 (01)00884838068445(21)77172 77074 (01)00884838068445(21)77074 77058 N/A 77114 N/A 77127 (01)00884838068445(21)77127 77160 (01)00884838068445(21)77160 77159 (01)00884838068445(21)77159 77182 (01)00884838068445(21)77182 77180 (01)00884838068445(21)77180 77171 (01)00884838068445(21)77171 77179 (01)00884838068445(21)77179 77048 N/A 77040 N/A 77034 N/A 77039 N/A 77035 N/A 77092 N/A 77140 (01)00884838068445(21)77140 77177 (01)00884838068445(21)77177 77181 (01)00884838068445(21)77181 77178 (01)00884838068445(21)77178 77041 N/A 77103 N/A 77193 (01)00884838068445(21)77193 77203 (01)00884838068445(21)77203 77134 (01)00884838068445(21)77134 77047 N/A 77051 N/A 77050 N/A 77060 N/A 77194 (01)00884838068445(21)77194 77111 N/A 77119 (01)00884838068445(21)77119 77167 (01)00884838068445(21)77167 77183 (01)00884838068445(21)77183 77185 (01)00884838068445(21)77185 77187 (01)00884838068445(21)77187 77188 (01)00884838068445(21)77188 77198 (01)00884838068445(21)77198 77195 (01)00884838068445(21)77195 77196 (01)00884838068445(21)77196 77199 N/A 77202 N/A 77037 N/A 77057 N/A 77072 N/A 77075 N/A 77084 N/A 77085 N/A 77078 N/A 77086 N/A 77101 N/A 77110 (01)00884838068445(21)77110 77038 (01)00884838068445(21)77038 77157 (01)00884838068445(21)77157 77156 (01)00884838068445(21)77156 77056 N/A 77052 N/A 77054 N/A 77053 N/A 77055 N/A 77064 N/A 77065 N/A 77067 N/A 77066 N/A 77068 N/A 77069 N/A 77105 N/A 77126 (01)00884838068445(21)77126 77125 (01)00884838068445(21)77125 77137 (01)00884838068445(21)77137 77139 (01)00884838068445(21)77139 77136 (01)00884838068445(21)77136 77135 (01)00884838068445(21)77135 77138 (01)00884838068445(21)77138 77148 (01)00884838068445(21)77148 77141 (01)00884838068445(21)77141 77030 N/A 77033 N/A 77036 N/A 77042 N/A 77044 N/A 77043 N/A 77032 N/A 77049 N/A 77061 N/A 77059 N/A 77071 N/A 77090 N/A 77088 N/A 77093 N/A 77091 N/A 77087 N/A 77097 N/A 77083 N/A 77106 N/A 77112 N/A 77107 (01)00884838068445(21)77107 77115 N/A 77100 (01)00884838068445(21)77100 77118 (01)00884838068445(21)77118 77121 (01)00884838068445(21)77121 77129 (01)00884838068445(21)77129 77120 (01)00884838068445(21)77120 77128 (01)00884838068445(21)77128 77151 (01)00884838068445(21)77151 77143 (01)00884838068445(21)77143 77158 (01)00884838068445(21)77158 77162 (01)00884838068445(21)77162 77161 (01)00884838068445(21)77161 77155 (01)00884838068445(21)77155 77190 (01)00884838068445(21)77190 77191 (01)00884838068445(21)77191 77150 (01)00884838068445(21)77150 77081 N/A 77117 (01)00884838068445(21)77117 77123 (01)00884838068445(21)77123 77166 (01)00884838068445(21)77166 77045 N/A 77063 N/A 77109 N/A 77201 (01)00884838068445(21)77201 77113 N/A 77147 (01)00884838068445(21)77147 77122 (01)00884838068445(21)77122 77154 (01)00884838068445(21)77154 77169 (01)00884838068445(21)77169 77168 (01)00884838068445(21)77168 77130 (01)00884838068445(21)77130 77094 N/A 77062 N/A
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAPhilips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 7283337 de octubre de 2026
- FDAAchieva 1.5T, Achieva 1.5T Conversion and Achieva 1.5T Initial system; Model Numbers (REF): 1. 781196 (Achieva30 de septiembre de 2026
- FDAIngenia 3.0T and Ingenia 3.0T CX; Model Numbers (REF): 1. 781342(3.0T), 2. 781377(3.0T), 3. 782103(3.0T), 4.30 de septiembre de 2026
- FDAIngenia 1.5T, Ingenia 1.5T CX and Ingenia 1.5T S; Model Numbers (REF): 1. 781315(1.5T), 2. 781341(1.5T), 3.30 de septiembre de 2026
- FDABanoful Top Orange Biscuit15 de septiembre de 2021
- FDASolero Generator US PG, Item No. H78712740000US0, Catalog No. 12740000US15 de septiembre de 2021
Preguntas sobre este retiro
¿Ingenia 1.5T CX-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body, or extremities Models: 781261 781262 fue retirado del mercado?
Sí. FDA publicó el retiro Z-2414-2021 el 15 de septiembre de 2021. Peligro reportado: When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter