Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2443-2012

Terumo Advanced Perfusion System 1 ASSEMBLY FLOW POD Product Usage: The Terumo Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.

Publicado el 3 de octubre de 2012 · Inicio del retiro: 10 de septiembre de 2012

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Terumo Cardiovascular Systems (CVS) received two reports of the false back flow alarms for the Terumo¿ Advanced Perfusion System 1 when there was no tubing in the flow sensor. Terumo CVS' investigation found that the voltage level on a circuit board was improperly set in a population of flow modules for Terumo System 1, the sensor operates correctly if the tubing is installed and it is highly unli

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Terumo Cardiovascular Systems Corporation
Distribuidor
Worldwide and US Nationwide Distribution
Lote
Catalog number: 802018 and serial numbers: 00032, 00042, 00056, 00147, 00623, 00700-00764, 00766-00820, and 00823-00847.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta