Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2476-2020

GlideScope Core OneTouch Smart Cable, REF: 0800 0601, an accessory to the GlideScope Video Laryngoscope System Per PLRA: The Core OneTouch Cables (Reusable Device) is ancillary equipment used with the Core 10 and Core 15 monitors. The monitor and workstation are intended to work with video endoscopes, in conjunction with ancillary equipment, for endoscopic procedures.

Publicado el 15 de julio de 2020 · Inicio del retiro: 5 de junio de 2020

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

When video laryngoscopy system users apply, twisting motions while connecting the blade in the HDMI port, or excessive torsion to the connection port, this force is transferred to the HDMI PCB, which may result in video synchronization interruption, loss of live image/blank screen, no cameras connected monitor warning, may present risk to patients, in which time to intubation is more critical.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Verathon, Inc.
Distribuidor
Worldwide Distribution: US (nationwide) and countries of: Australia and United Kingdom.
Lote
All serial numbers
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States