Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2604-2026

Omnipod DASH Pod. Includes the below Model/REF Numbers: 1. POD-BLE-I1-525 (5-Pack), POD-BLE-I1-520 (Single Pod). 2. POD-BLE-I1-529 (10-Pack), POD-BLE-I1-520 (Single Pod). 3. BLE-I1-529 (10-Pack), BLE-I1-520 (Single Pod). 4. BLE-C1-529 (10-Pack), BLE-C1-520 (Single Pod).

Publicado el 8 de julio de 2026 · Inicio del retiro: 20 de mayo de 2026

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.

Qué hacer

Voluntary: Firm initiated · Press Release

Detalles del retiro

Fabricante
Insulet Corporation
Distribuidor
Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Switzerland, Germany, Denmark, Finland, France, UK, Greece, Israel, Italy, Kuwait, Leichtenstein, Netherlands, Norway, Saudi Arabia, Sweden and Turkey.
Lote
1. Model/REF Number: POD-BLE-I1-525 (5-Pack), POD-BLE-I1-520 (Single Pod). Pack UDI-DI: 20385082000051. Single Pod UDI-DI: 10385082000023. Lot Numbers: PD1U11052421, PD1U11012421, PD1U11022421, PD1U02152521, PD1U02192521, PD1U04242522, PD1U10282522, PD1U10292521, PD1U10302521, PD1U10312521, PD1U11032521, PD1U11042521, PD1U01082621. 2. Model/REF Number: POD-BLE-I1-529 (10-Pack), POD-BLE-I1-520 (Single Pod). Pack UDI-DI: 20385082000020. Single Pod UDI-DI: 10385082000023. Lot Numbers: PD1U11192421, PD1U09102421, PD1U09092421, PD1U09062421, PD1U10242421, PD1U10252421, PD1U10282421, PD1U10292421, PD1U10312421, PD1U11042421. 3. Model/REF Number: BLE-I1-529 (10-Pack), BLE-I1-520 (Single Pod). Pack UDI-DI: 20385082000020. Single Pod UDI-DI: 10385082000023. Lot Numbers: PD1U11152421, PD1U11182421, PD1U11202421, PD1U09042421, PD1U11132421, PD1U11122421, PD1U11142421, PD1U11152411, PD1U11162421, PD1U04292521, PD1U09032421, PD1U11112421, PD1U04262521. 4. Model/REF Number: BLE-C1-529 (10-Pack), BLE-C1-520 (Single Pod). Pack UDI-DI: 20385082000112. Single Pod UDI-DI: 10385082000115. Lot Numbers: PD1U03042521, PD1U09052421, PD1U11072421, PD1U11082421, PD1U03062521, PD1U03032521, PD1U03072521, PD1U02282525, PD1U02282526, PD1U04252521, PD1U11062521, PD1U11072521, PD1U11102521, PD1U11112511, PD1U11112522, PD1U11122521.
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States