FDAFDA · Alimentos y medicamentos·N.º oficial Z-2707-2014

Spacelabs Medical Patient Monitor is a Physiological, Patient monitor With Arrhythmia Detection or Alarms. The Models are 91367, 91369, 91370, 91387, and 91393 are sold with Perioperative Mode (option D), which includes Start case/End case functionality, Upgrade Kit 040-1548-00. Spacelabs Healthcare patient monitors, functioning as either bedside or central monitors; passively display data generated by Spacelabs Healthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms. Key monitored parameters available on the models 91367, 91369, 91370 and 91387, when employing the Spacelabs Command Module, consist of ECG, respiration, invasive and noninvasive blood pressure, Sp02, temperature and cardiac output. Additional parameters and interfaces to other systems are also available depending on the parameter modules employed

Publicado el 1 de octubre de 2014 · Inicio del retiro: 26 de agosto de 2014

Resumen

FDA registra el retiro Z-2707-2014 de dispositivos médicos de Clase II, publicado el 1 de octubre de 2014 e iniciado el 26 de agosto de 2014. Empresa responsable: Spacelabs Healthcare Inc; Estado: Finalizado; Cantidad reportada: 82 units distributed in US and 140 units distributed outside US; Lote: The same KIT PART Number (PN 040-1458-00) identified on 12 units distributed IN the US and on 3 units distributed OUTSIDE of the US. These 15 units will receive serial numbers after the software upgrade is installed. THERE ARE 70 SERIAL NUMBERS ON UNITS DISTRIBUTED IN THE US: 040-1548-00, 1367-000033, 1369-001517, 1369-003051, 1369-003077, 1369-004247, 1369-005914, 1369-006877, 1369-007165, 1369-008519, 1369-010292, 1369-011637, 1369-017520, 1369-061217, 1370-000013, 1370-000014, 1370-000025, 1370-000095, 1370-000285, 1370-000449, 1370-000471, 1370-000579, 1370-000658, 1370-001700, 1370-002494, 1370-002510, 1370-003785, 1370-004176, 1370-004184, 1370-004275, 1370-004292, 1387-000465, 1387-000535, 1387-000543, 1387-001071, 1387-001677, 1387-006200, 1387-010824, 1387-010830, 1387-011914, 1387-014142, 1387-014159, 1387-014184, 1387-014186, 1387-014187, 1387-014188, 1387-014596, 1387-015721, 1387-015722, 1387-015723, 1387-016045, 1387-018424, 1387-018947, 1387-019766, 1387-019768, 1387-019769, 1387-019771, 1387-019772, 1387-100036, 1387-101646, 1387-104672, 1387-105801, 1387-105811, 1387-105812, 1387-105882, 1393-000442, 1393-000579, 1393-000580, 1393-000581, 1393-000582, and 387-002421. THERE ARE 137 SERIAL NUMBERS IDENTIFIED ON UNITS DISTRIBUTED OUTSIDE OF THE US: 1369-000011, 1369-000165, 1369-000407, 1369-000928, 1369-002976, 1369-003388, 1369-005316, 1369-005334, 1369-006100, 1369-006281, 1369-007338, 1369-009142, 1369-009143, 1369-009144, 1369-009145, 1369-009146, 1369-009889, 1369-009890, 1369-011741, 1369-011742, 1369-011743, 1369-011744, 1369-011745, 1369-013721, 1369-016648, 1369-016649, 1369-017759, 1369-017791, 1369-017792, 1370-000026, 1370-000725, 1370-000771, 1370-002373, 1370-002378, 1370-002380, 1370-002381, 1370-002382, 1370-002383, 1370-002384, 1370-003809, 1370-004133, 1370-004277, 1370-004509, 1370-004510, 1370-004511, 1370-004516, 1370-004617, 1370-200366, 1370-200370, 1370-200374, 1370-200375, 1370-200376, 1370-200378, 1370-200380, 1370-200471, 1370-200473, 1370-200474, 1370-200475, 1370-200476, 1370-200691, 1387-000044, 1387-000447, 1387-000467, 1387-000548, 1387-001086, 1387-002157, 1387-005006, 1387-008003, 1387-008310, 1387-010742, 1387-013241, 1387-013716, 1387-014610, 1387-018104, 1387-018105, 1387-018171, 1387-018176, 1387-018277, 1387-018638, 1387-018817, 1387-019563, 1387-100148, 1387-100150, 1387-100151, 1387-100155, 1387-100906, 1387-101061, 1387-101062, 1387-101063, 1387-101064, 1387-101065, 1387-101066, 1387-101067, 1387-102222, 1387-103550, 1387-103659, 1387-104462, 1387-105831, 1387-105832, 1387-105833, 1387-105834, 1393-000013, 1393-000517, 1393-000518, 1393-000519, 1393-000520, 1393-000521, 1393-000522, 1393-000523, 1393-000524, 1393-000525, 1393-000808, 1393-000809, 1393-000822, 1393-000831, 1393-000833, 1393-000834, 1393-000851, 1393-000852, 1393-00235, 1393-00236, 1393-00237, 1393-00238, 1393-00239, 1393-00240, 1393-00241, 1393-00335, 1393-00346, 1393-00347, 1393-00348, 1393-00349, 1393-00350, 1393-00351, 1393-00352, 1393-00353, 1393-00354, AND 1393-00355. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The clinician did not realize the monitor was in END CASE mode and was expecting audible alarms. Visual alarms continue to work normally. The firm performed a retrospective analysis of all recalls since 2011 and discovered 222 additional serial numbers were missed and were not included in the previous recall under RES 59456 (recall number Z-3032-2011).

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Spacelabs Healthcare Inc
Distribuidor
Worldwide Distribution- US (nationwide) including the states of: FL, GA, IA, KY, LA, MA, MI, MN, MO, MS, NJ, NY, OH, OK, PA, SC, TN, TX, WA, Guam and Puerto Rico and to the following countries: Argentina, Australia, Canada, China, Czech Republic, Finland, France, Germany, Greece, India, Israel, Italy, Netherlands, New Zealand, Panama, Paraguay, Poland, Singapore, Spain, Sri Lanka, Switzerland, Tunisia, United Kingdom, and Uruguay.
Lote
The same KIT PART Number (PN 040-1458-00) identified on 12 units distributed IN the US and on 3 units distributed OUTSIDE of the US. These 15 units will receive serial numbers after the software upgrade is installed. THERE ARE 70 SERIAL NUMBERS ON UNITS DISTRIBUTED IN THE US: 040-1548-00, 1367-000033, 1369-001517, 1369-003051, 1369-003077, 1369-004247, 1369-005914, 1369-006877, 1369-007165, 1369-008519, 1369-010292, 1369-011637, 1369-017520, 1369-061217, 1370-000013, 1370-000014, 1370-000025, 1370-000095, 1370-000285, 1370-000449, 1370-000471, 1370-000579, 1370-000658, 1370-001700, 1370-002494, 1370-002510, 1370-003785, 1370-004176, 1370-004184, 1370-004275, 1370-004292, 1387-000465, 1387-000535, 1387-000543, 1387-001071, 1387-001677, 1387-006200, 1387-010824, 1387-010830, 1387-011914, 1387-014142, 1387-014159, 1387-014184, 1387-014186, 1387-014187, 1387-014188, 1387-014596, 1387-015721, 1387-015722, 1387-015723, 1387-016045, 1387-018424, 1387-018947, 1387-019766, 1387-019768, 1387-019769, 1387-019771, 1387-019772, 1387-100036, 1387-101646, 1387-104672, 1387-105801, 1387-105811, 1387-105812, 1387-105882, 1393-000442, 1393-000579, 1393-000580, 1393-000581, 1393-000582, and 387-002421. THERE ARE 137 SERIAL NUMBERS IDENTIFIED ON UNITS DISTRIBUTED OUTSIDE OF THE US: 1369-000011, 1369-000165, 1369-000407, 1369-000928, 1369-002976, 1369-003388, 1369-005316, 1369-005334, 1369-006100, 1369-006281, 1369-007338, 1369-009142, 1369-009143, 1369-009144, 1369-009145, 1369-009146, 1369-009889, 1369-009890, 1369-011741, 1369-011742, 1369-011743, 1369-011744, 1369-011745, 1369-013721, 1369-016648, 1369-016649, 1369-017759, 1369-017791, 1369-017792, 1370-000026, 1370-000725, 1370-000771, 1370-002373, 1370-002378, 1370-002380, 1370-002381, 1370-002382, 1370-002383, 1370-002384, 1370-003809, 1370-004133, 1370-004277, 1370-004509, 1370-004510, 1370-004511, 1370-004516, 1370-004617, 1370-200366, 1370-200370, 1370-200374, 1370-200375, 1370-200376, 1370-200378, 1370-200380, 1370-200471, 1370-200473, 1370-200474, 1370-200475, 1370-200476, 1370-200691, 1387-000044, 1387-000447, 1387-000467, 1387-000548, 1387-001086, 1387-002157, 1387-005006, 1387-008003, 1387-008310, 1387-010742, 1387-013241, 1387-013716, 1387-014610, 1387-018104, 1387-018105, 1387-018171, 1387-018176, 1387-018277, 1387-018638, 1387-018817, 1387-019563, 1387-100148, 1387-100150, 1387-100151, 1387-100155, 1387-100906, 1387-101061, 1387-101062, 1387-101063, 1387-101064, 1387-101065, 1387-101066, 1387-101067, 1387-102222, 1387-103550, 1387-103659, 1387-104462, 1387-105831, 1387-105832, 1387-105833, 1387-105834, 1393-000013, 1393-000517, 1393-000518, 1393-000519, 1393-000520, 1393-000521, 1393-000522, 1393-000523, 1393-000524, 1393-000525, 1393-000808, 1393-000809, 1393-000822, 1393-000831, 1393-000833, 1393-000834, 1393-000851, 1393-000852, 1393-00235, 1393-00236, 1393-00237, 1393-00238, 1393-00239, 1393-00240, 1393-00241, 1393-00335, 1393-00346, 1393-00347, 1393-00348, 1393-00349, 1393-00350, 1393-00351, 1393-00352, 1393-00353, 1393-00354, AND 1393-00355.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Spacelabs Medical Patient Monitor is a Physiological, Patient monitor With Arrhythmia Detection or Alarms. The Models are 91367, 91369, 91370, 91387, and 91393 are sold with Perioperative Mode (option D), which includes Start case/End case functionality, Upgrade Kit 040-1548-00. Spacelabs Healthcare patient monitors, functioning as either bedside or central monitors; passively display data generated by Spacelabs Healthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms. Key monitored parameters available on the models 91367, 91369, 91370 and 91387, when employing the Spacelabs Command Module, consist of ECG, respiration, invasive and noninvasive blood pressure, Sp02, temperature and cardiac output. Additional parameters and interfaces to other systems are also available depending on the parameter modules employed fue retirado del mercado?

Sí. FDA publicó el retiro Z-2707-2014 el 1 de octubre de 2014. Peligro reportado: The clinician did not realize the monitor was in END CASE mode and was expecting audible alarms. Visual alarms continue to work normally. The firm performed a retrospective analysis of all recalls since 2011 and discovered 222 additional serial numbers were missed and were not included in the previous recall under RES 59456 (recall number Z-3032-2011).

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter