FDAFDA · Alimentos y medicamentos·N.º oficial Z-2707-2014
Spacelabs Medical Patient Monitor is a Physiological, Patient monitor With Arrhythmia Detection or Alarms. The Models are 91367, 91369, 91370, 91387, and 91393 are sold with Perioperative Mode (option D), which includes Start case/End case functionality, Upgrade Kit 040-1548-00. Spacelabs Healthcare patient monitors, functioning as either bedside or central monitors; passively display data generated by Spacelabs Healthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms. Key monitored parameters available on the models 91367, 91369, 91370 and 91387, when employing the Spacelabs Command Module, consist of ECG, respiration, invasive and noninvasive blood pressure, Sp02, temperature and cardiac output. Additional parameters and interfaces to other systems are also available depending on the parameter modules employed
Publicado el 1 de octubre de 2014 · Inicio del retiro: 26 de agosto de 2014
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The clinician did not realize the monitor was in END CASE mode and was expecting audible alarms. Visual alarms continue to work normally. The firm performed a retrospective analysis of all recalls since 2011 and discovered 222 additional serial numbers were missed and were not included in the previous recall under RES 59456 (recall number Z-3032-2011).
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Spacelabs Healthcare Inc
- Distribuidor
- Worldwide Distribution- US (nationwide) including the states of: FL, GA, IA, KY, LA, MA, MI, MN, MO, MS, NJ, NY, OH, OK, PA, SC, TN, TX, WA, Guam and Puerto Rico and to the following countries: Argentina, Australia, Canada, China, Czech Republic, Finland, France, Germany, Greece, India, Israel, Italy, Netherlands, New Zealand, Panama, Paraguay, Poland, Singapore, Spain, Sri Lanka, Switzerland, Tunisia, United Kingdom, and Uruguay.
- Lote
- The same KIT PART Number (PN 040-1458-00) identified on 12 units distributed IN the US and on 3 units distributed OUTSIDE of the US. These 15 units will receive serial numbers after the software upgrade is installed. THERE ARE 70 SERIAL NUMBERS ON UNITS DISTRIBUTED IN THE US: 040-1548-00, 1367-000033, 1369-001517, 1369-003051, 1369-003077, 1369-004247, 1369-005914, 1369-006877, 1369-007165, 1369-008519, 1369-010292, 1369-011637, 1369-017520, 1369-061217, 1370-000013, 1370-000014, 1370-000025, 1370-000095, 1370-000285, 1370-000449, 1370-000471, 1370-000579, 1370-000658, 1370-001700, 1370-002494, 1370-002510, 1370-003785, 1370-004176, 1370-004184, 1370-004275, 1370-004292, 1387-000465, 1387-000535, 1387-000543, 1387-001071, 1387-001677, 1387-006200, 1387-010824, 1387-010830, 1387-011914, 1387-014142, 1387-014159, 1387-014184, 1387-014186, 1387-014187, 1387-014188, 1387-014596, 1387-015721, 1387-015722, 1387-015723, 1387-016045, 1387-018424, 1387-018947, 1387-019766, 1387-019768, 1387-019769, 1387-019771, 1387-019772, 1387-100036, 1387-101646, 1387-104672, 1387-105801, 1387-105811, 1387-105812, 1387-105882, 1393-000442, 1393-000579, 1393-000580, 1393-000581, 1393-000582, and 387-002421. THERE ARE 137 SERIAL NUMBERS IDENTIFIED ON UNITS DISTRIBUTED OUTSIDE OF THE US: 1369-000011, 1369-000165, 1369-000407, 1369-000928, 1369-002976, 1369-003388, 1369-005316, 1369-005334, 1369-006100, 1369-006281, 1369-007338, 1369-009142, 1369-009143, 1369-009144, 1369-009145, 1369-009146, 1369-009889, 1369-009890, 1369-011741, 1369-011742, 1369-011743, 1369-011744, 1369-011745, 1369-013721, 1369-016648, 1369-016649, 1369-017759, 1369-017791, 1369-017792, 1370-000026, 1370-000725, 1370-000771, 1370-002373, 1370-002378, 1370-002380, 1370-002381, 1370-002382, 1370-002383, 1370-002384, 1370-003809, 1370-004133, 1370-004277, 1370-004509, 1370-004510, 1370-004511, 1370-004516, 1370-004617, 1370-200366, 1370-200370, 1370-200374, 1370-200375, 1370-200376, 1370-200378, 1370-200380, 1370-200471, 1370-200473, 1370-200474, 1370-200475, 1370-200476, 1370-200691, 1387-000044, 1387-000447, 1387-000467, 1387-000548, 1387-001086, 1387-002157, 1387-005006, 1387-008003, 1387-008310, 1387-010742, 1387-013241, 1387-013716, 1387-014610, 1387-018104, 1387-018105, 1387-018171, 1387-018176, 1387-018277, 1387-018638, 1387-018817, 1387-019563, 1387-100148, 1387-100150, 1387-100151, 1387-100155, 1387-100906, 1387-101061, 1387-101062, 1387-101063, 1387-101064, 1387-101065, 1387-101066, 1387-101067, 1387-102222, 1387-103550, 1387-103659, 1387-104462, 1387-105831, 1387-105832, 1387-105833, 1387-105834, 1393-000013, 1393-000517, 1393-000518, 1393-000519, 1393-000520, 1393-000521, 1393-000522, 1393-000523, 1393-000524, 1393-000525, 1393-000808, 1393-000809, 1393-000822, 1393-000831, 1393-000833, 1393-000834, 1393-000851, 1393-000852, 1393-00235, 1393-00236, 1393-00237, 1393-00238, 1393-00239, 1393-00240, 1393-00241, 1393-00335, 1393-00346, 1393-00347, 1393-00348, 1393-00349, 1393-00350, 1393-00351, 1393-00352, 1393-00353, 1393-00354, AND 1393-00355.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States