Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-3110-2026

Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser energy at wavelengths of 532 nm and 1064 nm through a fiber-optic handpiece. The system may be configured with an Intelligent Cooling Device (ICD) to provide epidermal cooling during laser treatment

Publicado el 16 de septiembre de 2026 · Inicio del retiro: 6 de agosto de 2026

Resumen

FDA registra el retiro Z-3110-2026 de dispositivos médicos de Clase II, publicado el 16 de septiembre de 2026 e iniciado el 6 de agosto de 2026. Empresa responsable: Lutronic Corporation; Estado: En curso; Cantidad reportada: 1065; Lote: Lot Code: Model: DermaV Software Version: 1.06 UDI #: 08800291820200 Serial Numbers: DV025A010, DV025A011, DV025A012, DV025A013, DV025A014, DV025A015 DV025A016, DV025A018, DV025A019, DV025A020, DV025A021, DV025A022, DV025B011, DV025B012, DV025B013, DV025B014, DV025B015, DV025D006, DV025D007, DV025D008, DV025D009, DV025D010, DV025E003, DV025E004, DV025E005, DV025E006, DV025E007, DV025E008, DV025E009, DV025E010, DV025E011, DV025E012, DV025E013, DV025E014, DV025E016, DV025E017, DV025F012, DV025F013, DV025F014, DV025F015, DV025F016, DV025G011, DV025G012, DV025G013, DV025G014, DV025G015, DV025G016, DV025G017, DV025G018, DV025G019, DV025G020, DV025J021, DV025J022, DV025J023, DV025J025, DV025J026, DV025J027, DV025J028, DV025J034, DV025J035, DV025J036, DV025J037, DV025J038, DV025J040, DV025K018, DV025K019, DV025K020, DV025K021, DV025K023, DV025K024, DV025K025, DV025K027, DV025M031, DV025M032, DV025M033, DV025M034, DV026B001, DV026B002, DV026B003, DV026B004, DV026B005, DV026B010, DV026B014, DV026B015, DV026B016, DV026B017, DV026B018, DV026C006, DV026C007, DV026C008, DV026C009, DV026C010, DV025M035. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn.

Qué hacer

Voluntary: Firm initiated · E-Mail

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Lutronic Corporation
Distribuidor
Worldwide distribution - US Nationwide and the countries of Canada and European Union.
Lote
Lot Code: Model: DermaV Software Version: 1.06 UDI #: 08800291820200 Serial Numbers: DV025A010, DV025A011, DV025A012, DV025A013, DV025A014, DV025A015 DV025A016, DV025A018, DV025A019, DV025A020, DV025A021, DV025A022, DV025B011, DV025B012, DV025B013, DV025B014, DV025B015, DV025D006, DV025D007, DV025D008, DV025D009, DV025D010, DV025E003, DV025E004, DV025E005, DV025E006, DV025E007, DV025E008, DV025E009, DV025E010, DV025E011, DV025E012, DV025E013, DV025E014, DV025E016, DV025E017, DV025F012, DV025F013, DV025F014, DV025F015, DV025F016, DV025G011, DV025G012, DV025G013, DV025G014, DV025G015, DV025G016, DV025G017, DV025G018, DV025G019, DV025G020, DV025J021, DV025J022, DV025J023, DV025J025, DV025J026, DV025J027, DV025J028, DV025J034, DV025J035, DV025J036, DV025J037, DV025J038, DV025J040, DV025K018, DV025K019, DV025K020, DV025K021, DV025K023, DV025K024, DV025K025, DV025K027, DV025M031, DV025M032, DV025M033, DV025M034, DV026B001, DV026B002, DV026B003, DV026B004, DV026B005, DV026B010, DV026B014, DV026B015, DV026B016, DV026B017, DV026B018, DV026C006, DV026C007, DV026C008, DV026C009, DV026C010, DV025M035
Clasificación de riesgo
Class II
Área (FDA)
device
País
Korea (the Republic of)

Retiros relacionados

Preguntas sobre este retiro

¿Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser energy at wavelengths of 532 nm and 1064 nm through a fiber-optic handpiece. The system may be configured with an Intelligent Cooling Device (ICD) to provide epidermal cooling during laser treatment fue retirado del mercado?

Sí. FDA publicó el retiro Z-3110-2026 el 16 de septiembre de 2026. Peligro reportado: Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · E-Mail