FDAFDA · Alimentos y medicamentos·N.º oficial Z-3332-2026

Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965 B96693 / 14987666542972 B12183 / 14987666540671 The AU480 Chemistry Analyzer measure analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material, and other accessories. This system is for in vitro diagnostic use only.

Publicado el 7 de octubre de 2026 · Inicio del retiro: 3 de agosto de 2026

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Beckman Coulter Mishima K.K.
Distribuidor
Worldwide - U.S. Nationwide distribution in the states of AR, CA, DE, FL, GA, GU, HI, IA, IL, IN, KS, KY, MN, MO, NC, NM, NY, OK, PA, SD, TX, and WI. The countries of Algeria, Australia, Belarus, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Cyprus, Czechia, Egypt, France, Germany, Guatemala, Hungary, India, Ireland, Italy, Japan, Kazakhstan, Korea, Republic of, Latvia, Malaysia, Morocco, Myanmar, Nepal, Nigeria, Pakistan, Poland, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, South Africa, Taiwan, Province of China, Turkey, United Arab Emirates, and Viet Nam.
Lote
Products containing affected software: Item Numbers / UDI-DI codes: Item Number: B96692 UDI-DI code: 14987666542965 Item Number: B96693 UDI-DI code: 14987666542972 Item Number: B12183 UDI-DI code: 14987666540671 Item Number: N3151400 UDI-DI code: Due to the age of the device and the fact that it has exceeded its storage period, the UDI number is not available.)
Clasificación de riesgo
Class II
Área (FDA)
device
País
Japan

Retiros relacionados

Preguntas sobre este retiro

¿Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965 B96693 / 14987666542972 B12183 / 14987666540671 The AU480 Chemistry Analyzer measure analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material, and other accessories. This system is for in vitro diagnostic use only. fue retirado del mercado?

Sí. FDA publicó el retiro Z-3332-2026 el 7 de octubre de 2026. Peligro reportado: Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter