Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0004-2013

Male, 16Fr, Straight-tip Intermittent Urinary Catheters M16. The product is distributed individually or in intermediate boxes containing 30 individual, sterile packaged catheters (PA/PE film and dialyzing paper pouches). There are 30 catheters packaged in sterile pouches that are placed into an intermediate box, 10 of these intermediate boxes packaged into one case. K072539 The Cure Catheter" device, M16, is an intermittent urinary catheter intended to be used by males for the purpose of bladder drainage. It is manufactured with conventional medical grade, latex-free, biocompatible materials. The straight-tip has been designed to eliminate trauma to the urethra and is provided in 16Fr size in an easy-to-open, sterile, single-use package. The coud®-tip catheter, M16C, is designed with a curved tip that makes it easier to pass through the curvature of the prostatic urethra in men. It is also used to pass through urethral strictures. This product is prescribed for people who have difficulty or are unable to pass a straight catheter through their urethra and into their bladder.

Publicado el 10 de octubre de 2012 · Inicio del retiro: 19 de septiembre de 2012

Resumen

FDA registra el retiro Z-0004-2013 de dispositivos médicos de Clase II, publicado el 10 de octubre de 2012 e iniciado el 19 de septiembre de 2012. Empresa responsable: Cure Medical LLC; Estado: Finalizado; Cantidad reportada: 420cases (= 4,200 boxes); Lote: Total affected product: M16, Lot 120425: 420 cases = 4,200 boxes = 126,000 each. 2017.03 is shown on the product labeling as the expiry date for the product, 4 years and 11 months from the date of packaging. The product, M16, is identified with the following UPC Codes for the individual pouches, intermediate boxes and case labels: ¿ Pouch Label UPC (M16) = *+M405M16 ¿ Box Label UPC (M16) = *+M405M163I* " Case Label UPC (M16) = *+M405M165K* Explanation of Lot Number: The lot number represents a planned packaging date of a set of work orders for various products, in the format YYMMDD at the contract manufacturers facility. At the order planning stage, the packaging date is estimated (based on production cycle time and pre-printed box delivery) and assigned to each work order The lot number represents a planned packaging date of a set of work orders for various products, in the format YYMMDD at the contract manufacturers facility. At the order planning stage, the packaging date is estimated (based on production cycle time and pre-printed box delivery) and assigned to each work order scheduled for packaging on this estimated date. Therefore, one lot number may be assigned to multiple catalog numbers. The lot number 120425 was assigned to both M16 and M16C products by the scheduling department Actual packaging of the M16 occurred from 23 May 2012 through 25 May 2012. Actual packaging of the M16C occurred from 19 May through 21 May 2012. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The firm initiated the product recall because labeling on some of the intermediate boxes of catheters identify the contents to be M16C, a coud®-tip catheters while the actual contents are M16Fr male, 16Fr, straight-tip catheters with the pouch and the case box labeled as M16Fr male, 16Fr, straight-tip catheters.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Cure Medical LLC
Distribuidor
Nationwide Distribution
Lote
Total affected product: M16, Lot 120425: 420 cases = 4,200 boxes = 126,000 each. 2017.03 is shown on the product labeling as the expiry date for the product, 4 years and 11 months from the date of packaging. The product, M16, is identified with the following UPC Codes for the individual pouches, intermediate boxes and case labels: ¿ Pouch Label UPC (M16) = *+M405M16 ¿ Box Label UPC (M16) = *+M405M163I* " Case Label UPC (M16) = *+M405M165K* Explanation of Lot Number: The lot number represents a planned packaging date of a set of work orders for various products, in the format YYMMDD at the contract manufacturers facility. At the order planning stage, the packaging date is estimated (based on production cycle time and pre-printed box delivery) and assigned to each work order The lot number represents a planned packaging date of a set of work orders for various products, in the format YYMMDD at the contract manufacturers facility. At the order planning stage, the packaging date is estimated (based on production cycle time and pre-printed box delivery) and assigned to each work order scheduled for packaging on this estimated date. Therefore, one lot number may be assigned to multiple catalog numbers. The lot number 120425 was assigned to both M16 and M16C products by the scheduling department Actual packaging of the M16 occurred from 23 May 2012 through 25 May 2012. Actual packaging of the M16C occurred from 19 May through 21 May 2012.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Male, 16Fr, Straight-tip Intermittent Urinary Catheters M16. The product is distributed individually or in intermediate boxes containing 30 individual, sterile packaged catheters (PA/PE film and dialyzing paper pouches). There are 30 catheters packaged in sterile pouches that are placed into an intermediate box, 10 of these intermediate boxes packaged into one case. K072539 The Cure Catheter" device, M16, is an intermittent urinary catheter intended to be used by males for the purpose of bladder drainage. It is manufactured with conventional medical grade, latex-free, biocompatible materials. The straight-tip has been designed to eliminate trauma to the urethra and is provided in 16Fr size in an easy-to-open, sterile, single-use package. The coud®-tip catheter, M16C, is designed with a curved tip that makes it easier to pass through the curvature of the prostatic urethra in men. It is also used to pass through urethral strictures. This product is prescribed for people who have difficulty or are unable to pass a straight catheter through their urethra and into their bladder. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0004-2013 el 10 de octubre de 2012. Peligro reportado: The firm initiated the product recall because labeling on some of the intermediate boxes of catheters identify the contents to be M16C, a coud®-tip catheters while the actual contents are M16Fr male, 16Fr, straight-tip catheters with the pouch and the case box labeled as M16Fr male, 16Fr, straight-tip catheters.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter