Merlin@Home Software Model EX2000 v8.2.2 for Merlin@Home Transmitter Models EX1150, EX1150W, EX1100, and EX1100W. The Merlin@home Transmitters are a component of St. Jude Medicals (SJM) Remote Care System. These Merlin@home transmitters are used to retrieve information from the SJM implantable medical device (IMD) using RF telemetry. The data is then sent back to the Merlin.net Patient Care Network through a built-in modem or, alternatively, through a USB cellular adapter or a USB based Broadband Kit. The transmitted data enables clinicians to remotely monitor a patients IMD using a secure, encrypted, password-protected web site, thus reducing the number of in-clinic visits a patient needs to make to their physician. The transmitter monitors the patients device daily between scheduled follow-ups and can alert a clinician if it detects anything out of the norm.
Publicado el 25 de octubre de 2017 · Inicio del retiro: 28 de agosto de 2017
Resumen
FDA registra el retiro Z-0033-2018 de dispositivos médicos de Clase II, publicado el 25 de octubre de 2017 e iniciado el 28 de agosto de 2017. Empresa responsable: St Jude Medical Inc.; Estado: Finalizado; Cantidad reportada: 495,011 units (391,093 in US); Lote: Model EX1100 - GTIN 05414734504782 and Model EX1150 - GTIN 05414734404334. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- St Jude Medical Inc.
- Distribuidor
- Worldwide Distribution
- Lote
- Model EX1100 - GTIN 05414734504782 and Model EX1150 - GTIN 05414734404334
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAELLIPSE VR, REF: CD1377-36QC Implantable cardioverter defibrillators (ICDs)14 de agosto de 2019
- FDAELLIPSE DR, REF: CD2377-36QC Implantable cardioverter defibrillators (ICDs)14 de agosto de 2019
- FDAELLIPSE DR, Tiered-therapy cardioverter/defibrillator, REF: CD2411-36C, UDI: 0541473450758514 de agosto de 2019
- FDAMerlin PCS programmer (Model # 3650), Software (Model # 3330)4 de julio de 2018
- FDAMerlin PCS 3650 Programmer with Model 3330 Software Version 23.1.1 The Merlin Patient Care System (PCS) Model25 de octubre de 2017
- FDAPM1214 NUANCE SR RF; PM1230 and PM2230 NUANCE MRI; PM2214 NUANCE DR RF These low voltage (LV) devices are25 de octubre de 2017
Preguntas sobre este retiro
¿Merlin@Home Software Model EX2000 v8.2.2 for Merlin@Home Transmitter Models EX1150, EX1150W, EX1100, and EX1100W. The Merlin@home Transmitters are a component of St. Jude Medicals (SJM) Remote Care System. These Merlin@home transmitters are used to retrieve information from the SJM implantable medical device (IMD) using RF telemetry. The data is then sent back to the Merlin.net Patient Care Network through a built-in modem or, alternatively, through a USB cellular adapter or a USB based Broadband Kit. The transmitted data enables clinicians to remotely monitor a patients IMD using a secure, encrypted, password-protected web site, thus reducing the number of in-clinic visits a patient needs to make to their physician. The transmitter monitors the patients device daily between scheduled follow-ups and can alert a clinician if it detects anything out of the norm. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0033-2018 el 25 de octubre de 2017. Peligro reportado: New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter