Sysmex XE-5000 Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060; catalog #063-7591-7 The Sysmex XE-5000 is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XE-5000 performs analyses using the following methods: RF/DC Detection Method, Sheath Row DC Detection Method, and Flow Cytometry Methods using a Semiconductor Laser. Particle characterization and identification is based on detection of forward scatter, fluorescence and adaptive cluster analysis. Using the same reagents as the XE-2100, the XE-5000 automatically classifies cells from blood and body fluids and carries out all processes automatically from aspiration of the sample to outputting the results.
Publicado el 14 de noviembre de 2012 · Inicio del retiro: 1 de junio de 2012
Resumen
FDA registra el retiro Z-0255-2013 de dispositivos médicos de Clase II, publicado el 14 de noviembre de 2012 e iniciado el 1 de junio de 2012. Empresa responsable: Sysmex America, Inc.; Estado: Finalizado; Cantidad reportada: 763 units; Lote: catalog #063-7591-7, all serial numbers. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Sysmex America, Inc.
- Distribuidor
- Nationwide distribution: USA including Puerto Rico.
- Lote
- catalog #063-7591-7, all serial numbers
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAThe CF-70 instrument (product code: LXG; Regulation Number: 862.2050) receives magazines containing stained22 de abril de 2020
- FDASysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-1022 de abril de 2020
- FDAThe UF-1000i is a fully automated urine particle analyzer intended for vitro diagnostic use in urinalysis. The19 de febrero de 2014
- FDASysmex XE-2100C Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe,14 de noviembre de 2012
- FDASysmex XE-2100 Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe,14 de noviembre de 2012
- FDASiemens Dimension Vista(R) Homocysteine Flex(R) Reagent Cartridge This is an in vitro diagnostic test for the14 de noviembre de 2012
Preguntas sobre este retiro
¿Sysmex XE-5000 Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060; catalog #063-7591-7 The Sysmex XE-5000 is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XE-5000 performs analyses using the following methods: RF/DC Detection Method, Sheath Row DC Detection Method, and Flow Cytometry Methods using a Semiconductor Laser. Particle characterization and identification is based on detection of forward scatter, fluorescence and adaptive cluster analysis. Using the same reagents as the XE-2100, the XE-5000 automatically classifies cells from blood and body fluids and carries out all processes automatically from aspiration of the sample to outputting the results. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0255-2013 el 14 de noviembre de 2012. Peligro reportado: There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter