IMMULITE/ IMMULITE 1000 Systems CMM CMV IgM Assay, REF/Catalog Number LKCM1(D) - 100 tests (US), Siemens Material Number (SMN) 10370301 (US), Unique Device Identification Number (UDI) 00630414985718 (US); and REF/Catalog Number LKCM1 - 100 tests (OUS); SMN 10381296 (OUS), UDI 00630414963747 (OUS); IVD --- Made in UK Siemens Healthcare Diagnostics Products Ltd., Glyn Rhonwy, Llanberis, Gwynedd, LL55 4EL UK For in vitro diagnostic use with the IMMULITE Systems Analyzers for the qualitative detection of IgM antibodies to Cytomegalovirus (CMV) in human serum or plasma (EDTA or heparinized), as an aid in the determination of an acute CMV infection.
Publicado el 2 de noviembre de 2016 · Inicio del retiro: 30 de agosto de 2016
Resumen
FDA registra el retiro Z-0319-2017 de dispositivos médicos de Clase II, publicado el 2 de noviembre de 2016 e iniciado el 30 de agosto de 2016. Empresa responsable: Siemens Healthcare Diagnostics, Inc.; Estado: Finalizado; Cantidad reportada: US: 5 kits; Foreign: 372 kits; Lote: UDI 00630414985718/Catalog # LKCM1(D): Kit lot D0330; UDI 00630414963747/Catalog # LKCM1: Kit lots 0330, 0331. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Multiple customers reported seeing an increase in imprecision with some patients resulting in high % CVs with the IMMULITE 2000/2000XPi CMV (IgM) assay starting with reagent lots 259 and above.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Siemens Healthcare Diagnostics, Inc.
- Distribuidor
- Worldwide distribution- US (Nationwide) and countries of: United Arab Emirates, Afghanistan, Angola, Argentina, Australia, Bangladesh, Belgium, Bulgaria, Bahrain, Bolivia, Brazil, Belarus, Switzerland, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Germany, Dominican Republic, Algeria, Ecuador, Egypt, Spain, France, Great Britain, Greece, Guatemala, Honduras, Hungary, India, Iraq, Italy, Jordan, Canary Islands, Kyrgyzstan, Rep. Korea (S), Kazakhstan, Lebanon, Lithuania, Luxembourg, Latvia, Macedonia, Mexico, Malaysia, Nicaragua, Netherlands, Norway, Nepal, Peru, Pakistan, Poland, Portugal, Paraguay, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, El Salvador, Syria, Turkmenistan, Tunisia, Turkey, Uruguay, Uzbekistan, Vietnam, Serbia,and South Africa.
- Lote
- UDI 00630414985718/Catalog # LKCM1(D): Kit lot D0330; UDI 00630414963747/Catalog # LKCM1: Kit lots 0330, 0331
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAAtellica CH Revised C-Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223;7 de octubre de 2026
- FDABeckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL26 de agosto de 2026
- FDASyva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mL26 de agosto de 2026
- FDASyva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219UL12 de agosto de 2026
- FDADesoximetasone Cream USP, 0.25% 100 g, Rx only, mfd. by Taro Pharmaceuticals Inc. Brampton, Ontario, Canada9 de noviembre de 2016
- FDAPentax Video Gastroscope is intended to be used with a Pentax video processor (including light source),2 de noviembre de 2016
Preguntas sobre este retiro
¿IMMULITE/ IMMULITE 1000 Systems CMM CMV IgM Assay, REF/Catalog Number LKCM1(D) - 100 tests (US), Siemens Material Number (SMN) 10370301 (US), Unique Device Identification Number (UDI) 00630414985718 (US); and REF/Catalog Number LKCM1 - 100 tests (OUS); SMN 10381296 (OUS), UDI 00630414963747 (OUS); IVD --- Made in UK Siemens Healthcare Diagnostics Products Ltd., Glyn Rhonwy, Llanberis, Gwynedd, LL55 4EL UK For in vitro diagnostic use with the IMMULITE Systems Analyzers for the qualitative detection of IgM antibodies to Cytomegalovirus (CMV) in human serum or plasma (EDTA or heparinized), as an aid in the determination of an acute CMV infection. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0319-2017 el 2 de noviembre de 2016. Peligro reportado: Multiple customers reported seeing an increase in imprecision with some patients resulting in high % CVs with the IMMULITE 2000/2000XPi CMV (IgM) assay starting with reagent lots 259 and above.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter