CAS I/II Absorbers that are parts of the BleaseSirius Anesthesia Workstation, the BleaseFocus Anesthesia Workstation with the following part numbers: 12200900; 12200901; 12200902; 12200903; 14200100, and service kits part number 050-0659-00 and 050-0901-00. BleaseFocus Anesthesia Workstation: The Blease Frontline Plus Range, Anesthesia Machines are intended for use in the hospital environment and locations not requiring portability. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of a mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. This device is intended for use only by a suitably qualified physician. BleaseSirius Anesthesia Workstation: The Spacelabs BleaseSirius Anesthesia Workstation is intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. The device is intended for use only by a suitably qualified physician.

Publicado el 27 de noviembre de 2013 · Inicio del retiro: 17 de octubre de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Spacelabs Healthcare Ltd. of United Kingdom recalls , and Service Kits, Part Number 050-0659-00 and 050-0901-00. The Bag to-Vent switch in CAS I/II Absorbers may fail due to loose fastening hardware in an absorber.

Qué hacer

Voluntary: Firm initiated · Telephone

Detalles del retiro

Fabricante
Del Mar Reynolds Medical, Ltd.
Distribuidor
Worldwide Distribution - USA (nationwide) including Oklahoma and North Carolina and Internationally to Libyan Arab Jamahiriya, Finland, Jordan, Mexico, Papua New Guinea, and Taiwan.
Lote
SERIAL NUMBERS IN THE US: ABSO-001754, ABSO-004081, ABSO-100111, ABSO-100112, ABSO-100113, ABSO-100114, ABSO-100117, ABSO-100120, ABSO-100121, ABSO-100122, ABSO-100123, ABSO-100124, ABSO-100125, ABSO-100126, ABSO-100127, ABSO-100128, ABSO-100132, ABSO-100134, and ABSO-100135. SERIAL NUMBERS INTERNATIONAL: ABSO-100098, ABSO-100099, abso-100100, ABSO-100101, abso-100102, ABSO-100103, abso-100104, abso-100105, abso-100106, ABSO-100107, ABSO-100108, ABSO-100109, ABSO-100110, ABSO-100129, ABSO-100130, ABSO-100131, ABSO-100136, AND ABSO-100137.
Clasificación de riesgo
Class I
Área (FDA)
device
País
United Kingdom