Leica BIOSYSTEMS; BOND Polymer Refine Red Detection; Catalog No: DS9390; Leica Biosystems Newcastle Ltd; Balliol Business Park West; Benton Lane; Newcastle Upon Tyne NE12 8EW; United Kingdom; www.LeicaBiosystems.com; Bond Polymer Refine Red Detection is a biotin-free, polymeric alkaline phosphate (AP)-linker antibody conjugate system for the detection of tissue-bound mouse and rabbit IgG and some mouse IgM primary antibodies. This detection system is for in vitro diagnostic use. Bond Polymer Refine Red Detection is a biotin-free, polymeric alkaline phosphatase (AP)-linker antibody conjugate system for the detection of tissue-bound mouse and rabbit IgG and some mouse IgM primary antibodies. It is intended for staining sections of formalin-fixed, paraffin-embedded tissue on the BOND automated system. The clinical interpretation of any staining or its absence should be complemented by morphological studies. Proper controls should be evaluated within the context of the patients clinical history and other diagnostic tests by a qualified pathologist. The Bond Polymer Refine Red Detection must be used with laboratory best practice in the use of tissue controls. For assurance, laboratories should stain each patient sample in conjunction with positive, negative and other tissue specific controls as needed.

Publicado el 22 de enero de 2014 · Inicio del retiro: 31 de julio de 2013

Resumen

FDA registra el retiro Z-0717-2014 de dispositivos médicos de Clase II, publicado el 22 de enero de 2014 e iniciado el 31 de julio de 2013. Empresa responsable: Leica Biosystems Newcastle Ltd.; Estado: Finalizado; Cantidad reportada: 1) 2,304 units; 2) 37 units; Lote: *** 1 ) Catalog Number: DS9390; Lot Numbers: 22017, 22139, 22262, 22337, 22411, 22514, 22541; *** 2) Catalog Number: DS9477; Lot Number: 22250. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The affected lots of the product may contain a contaminant of fungus that can cause a loss of staining sensitivity and/or obstructions to fluid delivery to the slide.

Qué hacer

Voluntary: Firm initiated · E-Mail

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Leica Biosystems Newcastle Ltd.
Distribuidor
Worldwide Distribution - USA including AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, and WV); and Internationally to Chile, Denmark, Finland, France, Germany, Ireland, Italy, Netherlands, Sweden, Switzerland, Poland, Romania, Singapore, South Korea, and the United Kingdom.
Lote
*** 1 ) Catalog Number: DS9390; Lot Numbers: 22017, 22139, 22262, 22337, 22411, 22514, 22541; *** 2) Catalog Number: DS9477; Lot Number: 22250
Clasificación de riesgo
Class II
Área (FDA)
device
País
United Kingdom

Retiros relacionados

Preguntas sobre este retiro

¿Leica BIOSYSTEMS; BOND Polymer Refine Red Detection; Catalog No: DS9390; Leica Biosystems Newcastle Ltd; Balliol Business Park West; Benton Lane; Newcastle Upon Tyne NE12 8EW; United Kingdom; www.LeicaBiosystems.com; Bond Polymer Refine Red Detection is a biotin-free, polymeric alkaline phosphate (AP)-linker antibody conjugate system for the detection of tissue-bound mouse and rabbit IgG and some mouse IgM primary antibodies. This detection system is for in vitro diagnostic use. Bond Polymer Refine Red Detection is a biotin-free, polymeric alkaline phosphatase (AP)-linker antibody conjugate system for the detection of tissue-bound mouse and rabbit IgG and some mouse IgM primary antibodies. It is intended for staining sections of formalin-fixed, paraffin-embedded tissue on the BOND automated system. The clinical interpretation of any staining or its absence should be complemented by morphological studies. Proper controls should be evaluated within the context of the patients clinical history and other diagnostic tests by a qualified pathologist. The Bond Polymer Refine Red Detection must be used with laboratory best practice in the use of tissue controls. For assurance, laboratories should stain each patient sample in conjunction with positive, negative and other tissue specific controls as needed. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0717-2014 el 22 de enero de 2014. Peligro reportado: The affected lots of the product may contain a contaminant of fungus that can cause a loss of staining sensitivity and/or obstructions to fluid delivery to the slide.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · E-Mail