FDAFDA · Alimentos y medicamentos·N.º oficial Z-0717-2014
Leica BIOSYSTEMS; BOND Polymer Refine Red Detection; Catalog No: DS9390; Leica Biosystems Newcastle Ltd; Balliol Business Park West; Benton Lane; Newcastle Upon Tyne NE12 8EW; United Kingdom; www.LeicaBiosystems.com; Bond Polymer Refine Red Detection is a biotin-free, polymeric alkaline phosphate (AP)-linker antibody conjugate system for the detection of tissue-bound mouse and rabbit IgG and some mouse IgM primary antibodies. This detection system is for in vitro diagnostic use. Bond Polymer Refine Red Detection is a biotin-free, polymeric alkaline phosphatase (AP)-linker antibody conjugate system for the detection of tissue-bound mouse and rabbit IgG and some mouse IgM primary antibodies. It is intended for staining sections of formalin-fixed, paraffin-embedded tissue on the BOND automated system. The clinical interpretation of any staining or its absence should be complemented by morphological studies. Proper controls should be evaluated within the context of the patients clinical history and other diagnostic tests by a qualified pathologist. The Bond Polymer Refine Red Detection must be used with laboratory best practice in the use of tissue controls. For assurance, laboratories should stain each patient sample in conjunction with positive, negative and other tissue specific controls as needed.
Publicado el 22 de enero de 2014 · Inicio del retiro: 31 de julio de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The affected lots of the product may contain a contaminant of fungus that can cause a loss of staining sensitivity and/or obstructions to fluid delivery to the slide.
Qué hacer
Voluntary: Firm initiated · E-Mail
Detalles del retiro
- Fabricante
- Leica Biosystems Newcastle Ltd.
- Distribuidor
- Worldwide Distribution - USA including AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, and WV); and Internationally to Chile, Denmark, Finland, France, Germany, Ireland, Italy, Netherlands, Sweden, Switzerland, Poland, Romania, Singapore, South Korea, and the United Kingdom.
- Lote
- *** 1 ) Catalog Number: DS9390; Lot Numbers: 22017, 22139, 22262, 22337, 22411, 22514, 22541; *** 2) Catalog Number: DS9477; Lot Number: 22250
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United Kingdom
¿Vendes productos como este?
Recalert vigila todo tu catálogo contra retiros como este, automáticamente.