BriteBlade Pro Single-Use Fiber Optic Mac 3, CE01120, Do Not Re-use, Rx Only, REF 040-713U, Sterile EO, Qty 10, (01)15055788723001 The device is intended for direct vision laryngoscopy to obtain an illuminated view of the vocal cords and the glottis by insertion into the oral cavity to manipulate the tongue and/or the epiglottis.

Publicado el 22 de enero de 2020 · Inicio del retiro: 27 de noviembre de 2019

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.

Qué hacer

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Detalles del retiro

Fabricante
Flexicare Medical Ltd.
Distribuidor
US: Nationwide OUS: Australia, Canada, China, Czech Republic, Denmark, Finland, Ireland, Italy, Kuwait, Lebanon, Netherlands, Panama, Romania, Russia, South Korea, UAE, UK
Lote
Part Number/Lot Number 040-713U 180700030 I CCW 040-713U 180900099 I CCW 040-713U 180901353 I CCW 040-713U 181000110 I CCW 040-713U 181000113 I CCW 040-713U 181100001 I CCW 040-713U 181100002 I CCW 040-713U 181200005 I CCW
Clasificación de riesgo
Class II
Área (FDA)
device
País
United Kingdom