BriteBlade Pro Single-se Fiber Optic Miller 2, CE 0120, Do not Re-use, Rx Only, REF, 040-722U, Sterile EO, Qty 10, (01) 15055788722783 The device is intended for direct vision laryngoscopy to obtain an illuminated view of the vocal cords and the glottis by insertion into the oral cavity to manipulate the tongue and/or the epiglottis.

Publicado el 22 de enero de 2020 · Inicio del retiro: 27 de noviembre de 2019

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.

Qué hacer

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Detalles del retiro

Fabricante
Flexicare Medical Ltd.
Distribuidor
US: Nationwide OUS: Australia, Canada, China, Czech Republic, Denmark, Finland, Ireland, Italy, Kuwait, Lebanon, Netherlands, Panama, Romania, Russia, South Korea, UAE, UK
Lote
Part Number/Lot Number 040-722U 181000115 I CCW 040-722U 181001520 I CCW 040-722U 181200006 I CCW 040-722U 181202055 I CCW 040-722U 190100006 I CCW
Clasificación de riesgo
Class II
Área (FDA)
device
País
United Kingdom