FDAFDA · Alimentos y medicamentos·N.º oficial Z-0846-2014

Simplexa Flu A/B & RSV Direct assay, Model MOL2650. The Focus Diagnostics Simplexa Flu A/B & RSV Direct assay is intended for use on the 3M Integrated Cycler instrument for the in vitro qualitative detection and differentiation of influenza A virus, influenza B virus, and respiratory syncytial virus (RSV) RNA in nasopharyngeal swabs (NPS) from human patients with signs and symptoms of respiratory tract infection in conjunction with clinical and epidemiological risk factors. This test is intended for use as an aid in the differential diagnosis of influenza A, influenza B, and RSV viral infections in humans and is not intended to detect influenza C.

Publicado el 5 de febrero de 2014 · Inicio del retiro: 10 de enero de 2014

Resumen

FDA registra el retiro Z-0846-2014 de dispositivos médicos de Clase II, publicado el 5 de febrero de 2014 e iniciado el 10 de enero de 2014. Empresa responsable: Focus Diagnostics Inc; Estado: Finalizado; Cantidad reportada: 1,052 kits; Lote: Model Number MOL2650 Lot Numbers: 24493, 24495, 24535, 24536. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Focus Diagnostics is initiating an urgent safety notice correction for Simplexa Flu A/B & RSV Direct assay (MOL2650) because Focus Diagnostics has received some customer complaints of Simplexa Flu A/B & RSV lots due to sporadic false positive RSV signals, which may result in a higher RSV false positive rate.

Qué hacer

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Focus Diagnostics Inc
Distribuidor
Worldwide Distribution - US (Nationwide) including states of: NY, MA, NJ, OH, CA, HI, CO, MN, FL, PA, IA, WI, AZ, TX, TN, IL, SD, MO, NM, AR, WV, MI and Internationally to: Sweden, Israel, Canada, and Slovenia. .
Lote
Model Number MOL2650 Lot Numbers: 24493, 24495, 24535, 24536.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Simplexa Flu A/B & RSV Direct assay, Model MOL2650. The Focus Diagnostics Simplexa Flu A/B & RSV Direct assay is intended for use on the 3M Integrated Cycler instrument for the in vitro qualitative detection and differentiation of influenza A virus, influenza B virus, and respiratory syncytial virus (RSV) RNA in nasopharyngeal swabs (NPS) from human patients with signs and symptoms of respiratory tract infection in conjunction with clinical and epidemiological risk factors. This test is intended for use as an aid in the differential diagnosis of influenza A, influenza B, and RSV viral infections in humans and is not intended to detect influenza C. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0846-2014 el 5 de febrero de 2014. Peligro reportado: Focus Diagnostics is initiating an urgent safety notice correction for Simplexa Flu A/B & RSV Direct assay (MOL2650) because Focus Diagnostics has received some customer complaints of Simplexa Flu A/B & RSV lots due to sporadic false positive RSV signals, which may result in a higher RSV false positive rate.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit