Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1205-2022

Proteus235- Proton Therapy System: to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation

Publicado el 8 de junio de 2022 · Inicio del retiro: 20 de abril de 2022

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

First layer of Uniform Scanning treatment fields is sometimes irradiated with incorrect Scanning Magnet setpoints resulting in the field size may be smaller than expected and the irradiated area may receive more dose than prescribed

Qué hacer

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Detalles del retiro

Fabricante
Ion Beam Applications S.A.
Distribuidor
Worldwide distribution - US Nationwide distribution in the states of IL, VA, WA and the country of Germany.
Lote
UDI-DI: (01)05404013801138 Serial Numbers: PAT108 (US), PAT112 (US), SAT116 (US). PAT107 (EU).
Clasificación de riesgo
Class II
Área (FDA)
device
País
Belgium

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta